A two-bath step test of a temperature sensor estimates the dynamic response of a temperature sensor by transferring it between two approximately uniform, stable baths at different temperatures. The apparatus consists of the two baths, independent reference thermometers, a reproducible immersion arrangement, and a recorder fast enough to resolve the temperature sensor response time. Rapid insertion into a liquid at a different temperature is a practical approximation to an instantaneous change in the sensor's surroundings.[ref]
Response Time and Temperature Sensors by Minco Staff. Second paragraph: rapid insertion into heated water or oil
Let the initial equilibrium be and the second bath be . Convert the recorded sensor output to temperature, then form the dimensionless residualFor a single lumped thermal model, , so a plot of against elapsed time is a straight line with slope . This provides a direct check on the assumed time constant; a curved plot suggests additional response stages or an imperfect input step. The conversion to temperature matters when sensor output is nonlinear.[ref]
Response Time and Temperature Sensors by Minco Staff. Paragraph beginning “For a rigorous approach”
For a constructed example, transfer a sensor from into a bath. Readings of , , and after , , and give residuals , , and . Each successive interval halves the residual, giving and .
Record the transfer delay, immersion depth, stirring conditions, probe sheath and readout filtering with the result. These affect heat transfer or the measured transient. A temperature sensor response time measured in one medium need not apply in another.[ref]
Guide on Secondary Thermometry: Industrial Platinum Resistance Thermometers by Jonathan Pearce and others. §4.3, printed p. 28: response depends on thermometer and surrounding medium
Articles by others on the same topic
There are currently no matching articles.