A two-channel covariance estimate of pulse-energy variance uses the component shared by paired readouts to estimate source fluctuations. In a stipulated linear model let and , with calibrated nonzero gains. If both noise terms are uncorrelated with and with each other,This follows by expanding the covariance. Correlation assumptions are essential measurement-model inputs; the uncertainty guide retains covariance terms explicitly.[ref]
For an original example, choose , and measured covariance . The inferred source variance is , giving a standard deviation of . Independent readout-noise variances and would give total variances and while preserving that cross-covariance.
But a shared noise covariance of raises the observed cross-covariance to . Ignoring it would infer a source standard deviation of , about too high.
A split-beam experiment records both channels for the same pulses, checks gains and measures dark or independent-input controls for correlated electronics. NIST's calibrated splitting apparatus supplies an example of the optical geometry, not a claim that it used this covariance estimator.[ref]
Calibration Service for Low-level Pulsed-Laser Radiometers at 1.06 µm: Pulse Energy and Peak Power by Rodney W. Leonhardt. §3.4, calibrated beam splitting
Matching pulse indices, bandwidth and integration windows is part of the proposed model. Articles by others on the same topic