I. H. McLaren coauthored with W. C. Wiley the 1955 Bendix study of time-of-flight mass spectrometry with improved mass resolving power. A notable part of their analysis was time-lag focusing: waiting between ion formation and application of an accelerating pulse lets initial velocities become position differences that can compensate some arrival-time spread.[ref]
A first-order calculation exposes the mechanism. Let be arrival time after immediate extraction for initial axial velocity and position . During a field-free delay , an ion moves to . Apart from the common delay added to all times,The partial derivatives are evaluated at and the reference position. First-order sensitivity to vanishes ifA positive delay requires opposite signs of these sensitivities. For example, and give . These values illustrate the cancellation, rather than reproducing an instrument setting from the paper.
This also reveals the tradeoff: the cancellation needs a nonzero position sensitivity, whereas ideal first-order space focusing sets . The paper's delayed-extraction condition and measured spectra therefore address competing sources of broadening, not a universal elimination of every initial-condition effect.[ref]
Time-of-Flight Mass Spectrometer with Improved Resolution by W. C. Wiley and I. H. McLaren. Printed p. 1154, equations 13–13a and discussion of the compromise between space resolution and time-lag energy focusing