I. H. McLaren coauthored with W. C. Wiley the 1955 Bendix study of time-of-flight mass spectrometry with improved mass resolving power. A notable part of their analysis was time-lag focusing: waiting between ion formation and application of an accelerating pulse lets initial velocities become position differences that can compensate some arrival-time spread.[ref]
A first-order calculation exposes the mechanism. Let be arrival time after immediate extraction for initial axial velocity and position . During a field-free delay , an ion moves to . Apart from the common delay added to all times,The partial derivatives are evaluated at and the reference position. First-order sensitivity to vanishes ifA positive delay requires opposite signs of these sensitivities. For example, and give . These values illustrate the cancellation, rather than reproducing an instrument setting from the paper.
This also reveals the tradeoff: the cancellation needs a nonzero position sensitivity, whereas ideal first-order space focusing sets . The paper's delayed-extraction condition and measured spectra therefore address competing sources of broadening, not a universal elimination of every initial-condition effect.[ref]
Time-of-Flight Mass Spectrometer with Improved Resolution by W. C. Wiley and I. H. McLaren. Printed p. 1154, equations 13–13a and discussion of the compromise between space resolution and time-lag energy focusing
Mass spectrometer 2026-11-05
A mass spectrometer is an instrument that produces or receives ions, separates them according to mass-to-charge ratio, and records their signals. Its major functions are supplied by an ion source, a mass analyzer and a mass spectrometer detector. The result is a mass spectrum; interpreting a peak also requires knowing which ionic chemical species produced it.
The 1955 instrument described by W. C. Wiley and I. H. McLaren provides a concrete apparatus example. An ion source and accelerating fields launch ions into an evacuated drift region toward a collector. Arrival times separate mass-to-charge ratios; the paper analyses how initial positions and velocities broaden the recorded peaks, and uses a two-field ion-gun geometry to improve focusing.[ref]
Time-of-Flight Mass Spectrometer with Improved Resolution by W. C. Wiley and I. H. McLaren. Page 1150, general description; subsequent resolution analysis and two-field ion gun
This is one analyzer design, not a definition requiring every mass spectrometer to measure flight time.For an ideal time-of-flight mass spectrometer, neglect initial speed and acceleration duration. Across an electric potential difference , an ion of positive electric charge gains . A field-free drift of length then givesIf one peak arrives at and another at , identical acceleration and drift conditions imply a mass-to-charge ratio ratio . It is a mass ratio only if the charge states are also equal. Peak position, width and intensity answer different questions: ionic identity, resolving ability and response-weighted abundance.
Time-of-Flight Mass Spectrometer with Improved Resolution by W. C. Wiley and I. H. McLaren 2026-11-05
By W. C. Wiley and I. H. McLaren. Review of Scientific Instruments 26(12), 1150–1157 (December 1955). DOI: doi.org/10.1063/1.1715212
Original paper: masspec.scripps.edu/learn/ms/pdf/1955_WileyWC.pdf
Time-of-flight mass spectrometry 2026-11-05
Time-of-flight mass spectrometry separates ions by measuring how long they take to reach a detector after a timed acceleration pulse. The simplest instrument has an ion source, an accelerating electric field, and a drift tube kept under vacuum. For ions of equal positive electric charge accelerated through the same electric potential difference, smaller mass gives greater speed and earlier arrival.[ref]
Time-of-Flight Mass Spectrometer with Improved Resolution by W. C. Wiley and I. H. McLaren. Review of Scientific Instruments 26, p. 1150, General Description
Let be the accelerating electric potential difference, the positive electric charge, the mass, and the field-free drift length. Neglecting initial speed and the time spent accelerating, conservation of energy givesFor an ideal drift at , a singly charged ion of mass arrives after approximately ; takes . Here is the unified atomic mass unit. Quadrupling mass doubles the ideal flight time. A doubly charged ion has the same mass-to-charge ratio as the singly charged ion and therefore the same ideal arrival time.
Real peaks broaden because ions begin at different positions and with different kinetic energies. In 1955, W. C. Wiley and I. H. McLaren analysed these effects and demonstrated an ion gun with two accelerating regions that reduced timing differences caused by initial position. Comparing calculated focusing conditions with measured spectra turned a timing idea into a practical improvement in mass spectrometry.[ref]
Time-of-Flight Mass Spectrometer with Improved Resolution by W. C. Wiley and I. H. McLaren. pp. 1150–1157; resolution analysis, two-field ion gun and experimental spectra
W. C. Wiley 2026-11-05
W. C. Wiley worked on time-of-flight mass spectrometry at Bendix Aviation Corporation Research Laboratories. With I. H. McLaren, Wiley published a 1955 instrument study addressing the arrival-time spread of ions with the same mass-to-charge ratio. Their apparatus combined an ion-producing source, accelerating electric fields, an evacuated drift region and a collector; its analysis separated broadening from initial position and initial motion.[ref]
Time-of-Flight Mass Spectrometer with Improved Resolution by W. C. Wiley and I. H. McLaren. Printed p. 1150, byline, affiliation and instrument description; pp. 1151–1154, space and initial-energy contributions
The basic timing principle can be derived without the focusing details. An initially stationary ion of positive electric charge accelerated through an electric potential difference acquires kinetic energy . Over a field-free drift length ,For a fixed charge state and instrument setting, differentiating gives the approximate mass resolving powerFor illustration, a arrival time with a timing width corresponds to , if that width is the relevant small peak-width measure.
The focusing problem addressed by Wiley and I. H. McLaren is therefore substantive: a spread of initial conditions can broaden a peak even when the detector timing is precise. Their work on space focusing and time-lag focusing changes the ion trajectories to reduce that broadening, rather than merely measuring the same broad peak with a faster clock.