Identifying flow-sensor gain and offset with timed transfers uses a known gas volume delivered over different durations to distinguish multiplicative response from accumulated zero error. In the model , a transfer of volume lasting producesThe observations form a linear regression with design matrix rows . If , gain and offset are identifiable only when the durations are not all equal. Repeating one duration measures the same combination of and more precisely, without separating them.
For two noise-free transfers,Stipulated 3 L transfers lasting 1 and 6 s, measured as 3.020 and 3.120 L, give and . The different durations expose an offset that one transfer could misinterpret as a gain error.
Timing spread and gain-offset estimator covariance quantifies the design. Suppose observations have independent equal volume-error variance , and define . Ordinary least squares givesUnder the stipulated model, wider timing separation improves offset precision. For the two durations above and L, the offset's standard uncertainty is about and the gain's about 0.00406.
The reference syringe volume was treated as known in those formulas. If it is uncertain, repeated sensor tests share that reference uncertainty. Indeed the observations identify only the product when both factors are unknown. Additional repetitions cannot establish an absolute gain without an independently characterized volume. The GUM treats fit uncertainties and common-reference correlations as distinct contributions.[ref]
Guide to the expression of uncertainty in measurement by JCGM. §4.2.5, fitted parameters; §5.2.4, common reference data
The primary spirometry statement's multiple-flow syringe checks test the practical response over different delivery rates.[ref]
Standardization of Spirometry 2019 Update by Brian L. Graham and others. Calibration and Quality Control, 3-L syringe over multiple flows
Actual leaks, flow-dependent nonlinearities and uncertain timing require a richer model than a constant gain and offset.Timing spread and gain-offset estimator covariance describes how delivery durations affect both precision and correlation when a known volume calibrates an integrated volumetric flow rate sensor. In the model , assume and durations are known and independent volume errors have common variance . Define .
The least squares slope estimates offset , while the intercept divided by estimates gain . Their covariance isThe negative covariance reflects a tradeoff: a larger fitted offset can be compensated by a smaller gain while preserving responses near the middle of the observed durations.
For an original design with L, durations 1 and 6 s, and L, the standard uncertainties are approximately 0.00283 L/s for offset and 0.00406 for gain. Their correlation is about . Moving the second duration to 2 s, with everything else stipulated unchanged, makes the offset uncertainty five times larger. Wider duration spread supplies more leverage for separating slope from intercept within this model.
Prediction shows why the covariance matters. For a new duration , the fitted mean response is , withAt the midpoint 3.5 s in the example, its standard uncertainty is L. Incorrectly dropping the gain-offset covariance would give about 0.0157 L. This is uncertainty of the fitted mean; an independent future observation adds its own error variance.
Centering time at makes the fitted midpoint response and slope uncorrelated under the stated equal-variance model. Transforming back to physical gain restores the covariance, so a change of coordinates does not remove shared uncertainty. The GUM explicitly retains such covariance terms in propagation.[ref]
Guide to the expression of uncertainty in measurement by JCGM. §5.2.2, equation 13, correlated-input propagation
Reference-volume uncertainty, uncertain timing and nonlinear volumetric flow rate response require additional terms beyond this original regression calculation. Articles by others on the same topic