A spirometer is an instrument for measuring gas volume changes associated with breathing. Its output supports spirometry. A displacement design follows the movement of a gas-containing chamber; a flow-measuring design records volumetric flow rate and integrates it. In a pneumotachograph, a measured pressure difference across a flow-resistive element supplies the signal used to estimate volumetric flow rate.
For an ideal piston with cross-sectional area and displacement , the transferred volume is . For a flow-measuring instrument,
These are different measurement chains for the same volume change. The 2019 primary technical statement explicitly discusses volume and flow instruments and their quality assurance.[ref]
An original error model makes integration drift visible. If a flow sensor reports , with constant gain and zero offset , its integrated result is
An offset of contributes 0.020 L during one second but 0.20 L during ten seconds. It can therefore change a long-expiration total more than an early timed volume, even though the flow offset is constant. A gain error instead scales both in proportion when no offset is present.
Spirometer calibration verification uses a known gas transfer to test the assembled measurement chain. The 2019 statement specifies at least daily verification with a 3-L syringe at several flows and a measured-volume accuracy requirement of ±3%.[ref] For a stipulated ideal 3.000 L transfer, a report of 3.090 L differs by 3%; identical total volumes at different transfer speeds can also expose a flow-dependent response.
Measurement calibration, leak checks and correct timing address different errors. Agreement with a syringe transfer establishes a specific instrument response; it does not by itself show that a person's breathing maneuver was complete or physiologically comparable to another attempt.
Spirometer calibration verification checks whether a spirometer measures a known input within specified limits. It differs from measurement calibration, which establishes the relation between input and indication, and from an adjustment that changes that relation. A satisfactory verification applies to the tested equipment, assembly and conditions; it does not validate a person's expiration maneuver.
The primary 2019 technical statement specifies daily verification with a 3-L syringe cycled over multiple flows. Its stated measured-volume accuracy requirement is ±3%, and a filter used for testing is also used during verification.[ref] The known volume is useful precisely because it can be delivered at different speeds.
An original gain-and-offset model explains this design. If measured flow is , integrating one transfer of known volume over duration gives
For the same at two durations,
A single transfer cannot distinguish gain from zero offset in this model; two different durations can.
Take stipulated 3.000 L transfers lasting 1 s and 6 s, reported as 3.020 and 3.120 L. The inferred offset is , while . The slow transfer is 4% high, exceeding the cited 3% limit. If both transfers instead read 3.060 L, this model gives and : a 2% gain error. These are constructed instrument responses, not a calibration record for an actual device.
Identifying flow-sensor gain and offset with timed transfers relies on the model assumptions. Nonlinear response, leaks or an inaccurately known syringe volume can produce other patterns. Reference-volume uncertainty belongs in the uncertainty budget, and repeated transfers share it rather than independently re-estimating it. A test at several flows is therefore more informative than repeating one convenient delivery speed.
Identifying flow-sensor gain and offset with timed transfers uses a known gas volume delivered over different durations to distinguish multiplicative response from accumulated zero error. In the model , a transfer of volume lasting produces
The observations form a linear regression with design matrix rows . If , gain and offset are identifiable only when the durations are not all equal. Repeating one duration measures the same combination of and more precisely, without separating them.
For two noise-free transfers,
Stipulated 3 L transfers lasting 1 and 6 s, measured as 3.020 and 3.120 L, give and . The different durations expose an offset that one transfer could misinterpret as a gain error.
Timing spread and gain-offset estimator covariance quantifies the design. Suppose observations have independent equal volume-error variance , and define . Ordinary least squares gives
Under the stipulated model, wider timing separation improves offset precision. For the two durations above and L, the offset's standard uncertainty is about and the gain's about 0.00406.
The reference syringe volume was treated as known in those formulas. If it is uncertain, repeated sensor tests share that reference uncertainty. Indeed the observations identify only the product when both factors are unknown. Additional repetitions cannot establish an absolute gain without an independently characterized volume. The GUM treats fit uncertainties and common-reference correlations as distinct contributions.[ref]
The primary spirometry statement's multiple-flow syringe checks test the practical response over different delivery rates.[ref] Actual leaks, flow-dependent nonlinearities and uncertain timing require a richer model than a constant gain and offset.
Timing spread and gain-offset estimator covariance describes how delivery durations affect both precision and correlation when a known volume calibrates an integrated volumetric flow rate sensor. In the model , assume and durations are known and independent volume errors have common variance . Define .
The least squares slope estimates offset , while the intercept divided by estimates gain . Their covariance is
The negative covariance reflects a tradeoff: a larger fitted offset can be compensated by a smaller gain while preserving responses near the middle of the observed durations.
For an original design with L, durations 1 and 6 s, and L, the standard uncertainties are approximately 0.00283 L/s for offset and 0.00406 for gain. Their correlation is about . Moving the second duration to 2 s, with everything else stipulated unchanged, makes the offset uncertainty five times larger. Wider duration spread supplies more leverage for separating slope from intercept within this model.
Prediction shows why the covariance matters. For a new duration , the fitted mean response is , with
At the midpoint 3.5 s in the example, its standard uncertainty is L. Incorrectly dropping the gain-offset covariance would give about 0.0157 L. This is uncertainty of the fitted mean; an independent future observation adds its own error variance.
Centering time at makes the fitted midpoint response and slope uncorrelated under the stated equal-variance model. Transforming back to physical gain restores the covariance, so a change of coordinates does not remove shared uncertainty. The GUM explicitly retains such covariance terms in propagation.[ref] Reference-volume uncertainty, uncertain timing and nonlinear volumetric flow rate response require additional terms beyond this original regression calculation.