Mix an unknown single-frequency laser with a nearby comb tooth on a photodiode. Define the signed beat , so
The beat arises from the cross term in light intensity:
An electronic detector can follow this slow difference even though it cannot resolve either optical oscillation. The square-law expression also shows why a real-valued intensity trace by itself loses the sign of . A small known upward shift of the chosen tooth decreases when the unknown lies above it and increases when it lies below, provided the tooth assignment remains unchanged.
For an original example, stipulate
Then and . Choosing the opposite sign would give instead. The stipulated optical frequency is countable because the large optical value is expressed using the integer and three electronic frequencies; the diagram's two controls establish the comb, and this additional beat places the unknown against it.
Both comb readouts use one calibrated microwave reference. Feedback fixes and , determining each tooth once is identified. A frequency counter measures the slow beat, transferring the calibration to the optical frequency.[ref]
One beat magnitude does not tell us the integer . In the example, every proposed optical frequency produces the same signed beat with its corresponding tooth. A coarse independent optical measurement can select the integer: once the beat sign is known, an error bound below suffices if the candidate is selected by nearest distance.
There is also a useful two-spacing check. Keep the unknown laser and fixed, track the same tooth continuously, and increase by a stipulated . Its frequency rises by , so the signed beat falls from to . Hence
This inference fails if the unknown drifts appreciably, the offset changes unnoticed, the beat sign is lost, or the tracker jumps to another tooth. The assumed tooth motion stays well below half the spacing in this example. If also changes by a measured amount, use .
For a fixed verified tooth number, the first-order frequency error is . Errors in these readings can be correlated because they share a reference. A small beat uncertainty therefore does not imply the same absolute optical accuracy: the reference's calibration and the factor remain in the error budget. A wrong tooth number is a separate discrete error, not a small correction. The beauty of the comb is that two controlled parameters organize millions of optical reference points, while the unknown's position is read electronically.