The closest components above differ by . At , their separation is about . A visible spectral line near would consequently require spectral resolving power of order . An ordinary low-resolution spectrometer would merge the components.
A suitable arrangement places an excited low-pressure strontium discharge or a transversely observed atomic beam between wide electromagnet pole pieces. Collect the emission perpendicular to the magnetic field, select the desired spectral line with a prefilter or monochromator, and send it through a high-finesse scanning Fabry–Pérot interferometer to a detector. A rotatable linear polarizer separates the two components from the four components. Calibrate both the interferometer's free spectral range and the local magnetic field; take zero-field and several field-strength scans, and fit component positions against .
The magnetic field should be uniform across the viewed volume, with its associated Zeeman splitting larger than the combined natural linewidth, Doppler broadening, pressure broadening and instrumental width, but much smaller than the separations of the relevant atomic fine-structure levels. A collimated transverse atomic beam, low collision rate and restricted collection angle reduce unwanted spectral line broadening. Avoid appreciable power broadening, resolve or model isotope and hyperfine structure, and check for self-absorption. A zero-field scan and the predicted linear scaling help distinguish actual Zeeman splitting components from neighbouring spectral lines.