The Hall-plateau voltmeter calibration is obtained as follows. Let be a plateau height in displayed units after any voltage offset is removed, with identified integer plateau index . The true Hall voltage is , so the calibration factor would be
The nominal current is . If several plateau heights are available, their inverse-integer ratios help identify indices and a fit of against can separate a constant offset from the gain. Reversing magnetic field and antisymmetrizing the voltage also removes an even offset. A unique absolute index should not be asserted solely from an unspecified sequence of steps.
For an established exact integer and independent current and height errors,
The current alone contributes relative uncertainty; it is a common calibration uncertainty and does not average away by fitting more plateaux. Voltage-offset uncertainty and deviations from quantization must also be included where relevant. The width cancels because the Hall voltage divided by current is a sheet Hall resistance; it would enter the transverse electric field, not this calibration.
TODO — missing source data: the held PDF's last-page stream refers to an absent /Im6 graphic, and its recovery log reports an unknown XObject. The supplied TeX records the same source gap. No plateau heights, labels or axis values survive, so neither a numerical nor a unique can be read. The numerical voltmeter calibration and its complete graph-reading error cannot be supplied without that graph; the formula and known current contribution above are determined.