Independent fluctuations add as variances. A useful parametrization of the electromagnetic calorimeter energy resolution is
The convention makes dimensionless when is inserted in GeV. The three terms describe different mechanisms.
The stochastic term comes from fluctuations in the electromagnetic shower, scintillation production and detected photoelectron counts. If the number of detected photoelectrons is , Poisson distribution gives , hence . A sampling electromagnetic calorimeter also has fluctuations in the fraction deposited in its active material, contributing to the same approximate scaling. A homogeneous detector suppresses that sampling contribution.
The noise term represents approximately fixed absolute electronics and dark-current noise for a chosen reconstruction window. Additional collision pileup fluctuations can behave similarly over a specified operating range. A nearly constant gives .
The constant term describes residual fractional cell-calibration errors, gain drifts, temperature dependence, nonuniform light collection, dead material, and imperfect correction for longitudinal and transverse leakage. These effects often scale approximately with the incident energy, so they approach a constant fractional error. For electrons, upstream bremsstrahlung and conversion of the radiated photons spread the deposit over multiple cells; material mapping and clustering are essential. Leakage and material losses can also produce non-Gaussian tails, so the three-term parametrization describes a resolution core rather than every feature of the response. Noise dominates at low energy, stochastic fluctuations at intermediate energy, and the constant term eventually limits performance.

Articles by others on the same topic (0)

There are currently no matching articles.