Pulse generator 2026-11-05
A pulse generator produces controlled electrical pulses for exciting, triggering or testing an electrical circuit. The useful specification includes high and low levels, pulse width, repetition period, edge behavior and the timing relationship to a trigger. A repeating pulse train and a single triggered pulse impose different test conditions even when their individual pulse shapes agree.
For an original ideal train with low level zero, high level , pulse width and period , define duty cycle and repetition frequency . Direct averaging givesFor a resistive load , the energy per pulse and mean active power areMean potential difference and RMS potential difference are different summaries; squaring the mean would underestimate heating for a nonconstant waveform.
Take actually delivered into , and . Then , , mean potential difference is , and RMS potential difference is about . The pulse deposits and average load power is . These are stipulated rectangular-pulse values, not instrument limits.
Real pulse width needs a threshold definition. Keysight's primary control documentation measures it between the rising and falling 50% crossings and provides duty-cycle entry as an alternative.[ref]
Pulse Width in Keysight BenchVue Function Generator documentation. Pulse Width, definition and duty-cycle setting
Rise time, edge asymmetry and trigger timing must be measured or specified separately; a nominal width alone does not describe the full waveform.Termination also changes what reaches the load. The existing Keysight documentation explains a fixed series output impedance and the difference between a selected load setting and the actual connected load. Its example gives twice the displayed amplitude at an open circuit when the display assumes a matched 50-ohm load.[ref]
Output Load Termination in the Keysight BenchVue Function Generator documentation. Output Load Termination, first paragraph and open-circuit example
Use the delivered pulse levels in the energy calculation.A proposed test connects the intended termination, records the pulse at the load with an oscilloscope, and checks levels, crossing times and repetition intervals. Varying pulse width at fixed amplitude and repetition rate should change resistive average power in proportion to under the ideal model. Loading and finite edge durations explain departures that should not automatically be assigned to the programmed timing.
Pulse width 2026-11-05
Pulse width is the time separation of specified crossings on a pulse's leading and trailing edges. The threshold convention is essential when the edges are not instantaneous. Keysight's primary pulse-generator documentation uses the rising and next falling 50-percent crossings, and allows a duty-cycle setting as an alternative control.[ref]
Pulse Width in Keysight BenchVue Function Generator documentation. Pulse Width, 50-percent definition and duty-cycle setting
For low and high levels and , define the fractional threshold as with . A positive pulse's width at that level is . The half-height convention is related to full width at half maximum for an isolated positive pulse above a defined baseline; it is not necessarily the duration of the flat top.
For an original asymmetric trapezoid, let a full linear rise take , the plateau last , and a full linear fall take . These ramp durations refer to the full low-to-high excursions, not the commonly quoted 10–90 percent rise interval. The crossing times giveThe 50-percent width is , while the first departure from the low level to final return lasts . At a 90-percent threshold the width is only . All describe the same waveform under different conventions.
For a repeating pulse train, duty cycle compares the specified width to the repetition period. Finite ramps mean that width times peak amplitude alone need not give the waveform integral or energy. Electrical loading and bandwidth also change the delivered crossings; the pulse generator setting must be checked at the actual load.
A measurement records both edges on an oscilloscope with a common level definition. Pulse-width uncertainty from correlated edge timing depends on , including their covariance. A common trigger-time translation can cancel from width while remaining in absolute arrival time. Ringing with several crossings, hysteresis, negative pulses, or multiple peaks needs an explicit crossing-selection convention rather than an unspecified single duration.