Centrifugation 2026-10-13
Centrifugation separates suspended material by its motion relative to a rotating liquid. A centrifuge rotor creates an outward centrifugal acceleration in the rotating reference frame; material denser than its surroundings tends to move outward, opposed by drag force. Collected material can form a centrifugation pellet, leaving a supernatant, or remain in separated zones within a density gradient. The useful question is which physical difference the separation exploits: a difference in sedimentation coefficient or a difference in buoyant density.
Figure 1. . Original schematic of one swung-out tube during rotation and a conceptual pelleting comparison. Rotor radius increases toward the tube's outer end. Colors and particle counts illustrate incomplete separation and trapping; they are not measured recovery fractions. The tube remains horizontal in this depiction of the rotating geometry. Details.
Centrifugation makes otherwise slow settling useful for studying biological cells, organelles, proteins and DNA. Separation and identification are different steps: a centrifugation pellet is a collected fraction, not proof that it contains only one kind of organelle. The lysozyme wall-removal experiment illustrates a complementary readout: testing material released into the separated liquid can be more informative than inspecting the retained solids.
Centrifuge rotor 2026-10-13
A centrifuge rotor holds sample vessels around the rotation axis. A fixed-angle centrifuge rotor holds tubes at a fixed inclination; a swinging-bucket centrifuge rotor lets them swing outward during rotation. Material moves away from the axis, so a centrifugation pellet can first collect against an outer tube wall rather than travel along the entire tube axis. Geometry sets both the rotor radius and the travel distance.[ref]
Relative centrifugal force is the dimensionless ratio of the local centrifugal acceleration magnitude to the conventional gravitational reference . Despite its name, it is an acceleration ratio: the actual force on a specified mass is . With angular speed and rotor radius , . If is the numerical rotational speed in revolutions per minute and the numerical rotor radius in centimetres, unit conversion gives
Thus rpm alone does not specify the centrifugal acceleration.[ref]
Figure 1.
Relative centrifugal force for stipulated rotor radii and rotational speeds
. Original theoretical curves calculated from with . The rotor radii of , and are chosen examples, not specifications of real centrifuge rotors. The plotted range is not an instrument operating limit or a recommendation. Details.
For a stipulated , rotor radii of and give approximately and times . Doubling the rotor radius doubles relative centrifugal force; doubling rpm quadruples it. Even equal relative centrifugal forces do not guarantee equal separation: travel distance, elapsed time, viscosity, sample density and particle shape still matter. These are comparisons of a mathematical model, not instructions for running an instrument.
Rotor radius 2026-10-13
The rotor radius is the perpendicular distance from the rotation axis to a specified sample position, not the tube's length. A filled tube spans a range . Consequently, one rotational speed produces different local centrifugal accelerations along that range. A reported relative centrifugal force needs a stated reference position, such as the outermost sample radius.