Tetrazolium contrast uncertainty quantifies uncertainty in the four-reading cell-free tetrazolium interaction contrast, including correlations between its inputs. In the order reagent/test, no-reagent/test, reagent/vehicle and no-reagent/vehicle, write
For finite input variances and known covariance matrix , the variance is exactly
The exactness here follows from linearity of the contrast. JCGM's general uncertainty propagation framework uses sensitivity coefficients and correlated input quantities; this contrast has coefficients independent of the readings.[ref]
An original example gives each reading an independent residual with standard uncertainty and all four readings the same additive offset with standard uncertainty . Then
Since , the common offset cancels, giving . Treating those four appearances of one shared offset as independent would instead give . The distinction concerns the physical source of variation, not the number of numbers in the subtraction.
A shared multiplicative gain does not cancel. If , with independent zero-mean gain error , common offset and independent residuals , then
For chosen , and , the combined standard uncertainty is approximately . The gain contribution is a relative uncertainty, whereas the residual contribution is in absorbance units.
The input model should distinguish repeated readings of one preparation from independently prepared controls. A small propagated uncertainty does not test whether controls reproduce the sample's chemistry or whether a nonzero optical contrast identifies a particular formazan. Those are separate questions about the measurement model.