Mass analyzer 2026-11-05
A mass analyzer is the part of a mass spectrometer that makes ions distinguishable or selectively transmits them according to mass-to-charge ratio. The measured coordinate can be position, arrival time or transmission under a specified field setting. A separate mass spectrometer detector records the resulting signal.
A quadrupole mass analyzer selects trajectories using combined static and time-varying electric fields: chosen mass-to-charge values remain on stable paths while others are lost from the transmitted beam.[ref] A time-of-flight mass spectrometer instead encodes the ratio in arrival time after acceleration. Different physical designs can therefore provide the same kind of ionic-mass information through different observables.
An original magnetic example exposes an important assumption. For speed perpendicular to a uniform magnetic field , balancing the Lorentz force and circular acceleration givesIf the incoming ions have the same selected speed and charge, mass values in the ratio give bending radii in that ratio, about . But if ions instead start negligibly slowly and gain energy from the same electric potential difference , substituting gives . Their radius ratio is then . The initial-condition choice changes the mapping.
Analysis requires this mapping and its assumptions, not just separated peaks. Mass resolving power describes the ability to distinguish nearby peak coordinates under a stated width criterion; accurate assignment of those coordinates additionally needs instrument calibration and the correct charge-state interpretation.
Mass spectrometer 2026-11-05
A mass spectrometer is an instrument that produces or receives ions, separates them according to mass-to-charge ratio, and records their signals. Its major functions are supplied by an ion source, a mass analyzer and a mass spectrometer detector. The result is a mass spectrum; interpreting a peak also requires knowing which ionic chemical species produced it.
The 1955 instrument described by W. C. Wiley and I. H. McLaren provides a concrete apparatus example. An ion source and accelerating fields launch ions into an evacuated drift region toward a collector. Arrival times separate mass-to-charge ratios; the paper analyses how initial positions and velocities broaden the recorded peaks, and uses a two-field ion-gun geometry to improve focusing.[ref]
Time-of-Flight Mass Spectrometer with Improved Resolution by W. C. Wiley and I. H. McLaren. Page 1150, general description; subsequent resolution analysis and two-field ion gun
This is one analyzer design, not a definition requiring every mass spectrometer to measure flight time.For an ideal time-of-flight mass spectrometer, neglect initial speed and acceleration duration. Across an electric potential difference , an ion of positive electric charge gains . A field-free drift of length then givesIf one peak arrives at and another at , identical acceleration and drift conditions imply a mass-to-charge ratio ratio . It is a mass ratio only if the charge states are also equal. Peak position, width and intensity answer different questions: ionic identity, resolving ability and response-weighted abundance.
Time-of-flight mass spectrometer 2026-11-05
A time-of-flight mass spectrometer is a mass spectrometer whose analyzer infers mass-to-charge ratio from ion arrival times. A simple arrangement uses an ion source, pulsed acceleration, a drift region under vacuum and a mass spectrometer detector. The 1955 Wiley–McLaren instrument connected such an apparatus to calculations of how initial position and velocity spread affect resolution.[ref]
Time-of-Flight Mass Spectrometer with Improved Resolution by W. C. Wiley and I. H. McLaren. Page 1150, general description; two-field focusing and resolution analysis
In an ideal drift, for flight length , positive charge and accelerating electric potential difference . Recorded arrival time can also include a time-of-flight timing offset . A useful ideal instrument calibration model iswhere is the numerical mass-to-charge coordinate in a declared unit and absorbs the instrument geometry and unit conversion.
For original invented standards, let arrive at and at . Since the square roots are 5 and 10, and . An unknown arriving at then has . Ignoring the offset and scaling from the first standard alone would instead give . A timing offset becomes a mass-assignment error when omitted from the model.
Real peaks have finite widths. Space focusing and time-lag focusing address specified sources of that spread; neither is simply a longer wait before detection. Calibration standards, acquisition timing and focusing conditions therefore belong in the interpretation of a measured spectrum.
Time-of-flight timing offset 2026-11-05
A time-of-flight timing offset, , is an additive time-origin parameter in a model connecting recorded ion arrival time with physical flight time. In a simple time-of-flight mass spectrometer calibration,where is the numerical mass-to-charge ratio in declared units. A 2017 instrument-analysis paper uses this equation and determines instrument-dependent constants from identified isotope peaks.[ref]
An original error calculation shows why the offset matters. Suppose the physical flight contribution is and , so the recorded time is . With held fixed, ignoring the offset multiplies the inferred coordinate bya overestimate. This deliberately large example isolates the effect; it does not describe a measured instrument error.
With fixed, a small error in the fitted offset propagates approximately asAn offset uncertainty of at a corrected flight time of therefore contributes about relative uncertainty in this approximation. The same absolute offset error has greater relative impact on earlier arrivals.
This parameter concerns timing, whereas time-lag focusing is a physical delayed-extraction technique affecting the ion packet. Fitting an offset does not by itself sharpen a peak. More detailed instruments can require additional calibration terms; the stated equation is a model, not a universal transfer function.