Temperature sensor response time
ID: temperature-sensor-response-time
A temperature sensor response time specifies how long a temperature sensor takes to complete a stated fraction of its response to a change in the surrounding temperature. The fraction and the test conditions are part of the specification: a 90% response time and a 63.2% response time describe different points on the same record. Rapid immersion in a different-temperature liquid approximates the required step input.[ref]
Response Time and Temperature Sensors by Minco Staff. Opening two paragraphs: response criterion and immersion test
For a lumped thermal model with sensor heat capacity , constant thermal conductance to a bath at , and negligible internally generated heat, the energy conservation equation and its solution areThus the time constant is the time to complete of the change. The time to complete a fraction is , so , whereas the elapsed time between 10% and 90% is . For example, gives and .
A two-bath step test of a temperature sensor tests this exponential function model rather than assuming it. Several internal heat capacity elements, a protective sheath, or electronic filtering can produce several response stages. The thermometry guide reports fast and slow components in tested platinum thermometer assemblies; a single quoted time constant need not describe their entire approach to equilibrium.[ref]