Capillary electrophoresis separates species by their motion in an electric field inside a narrow liquid-filled tube. In its free-solution form, different electrophoretic mobilities produce different drift speeds; a detector at a fixed position records the arriving bands as an electropherogram. The liquid itself can also move by electroosmotic flow, so arrival order is not determined by molecular charge alone.
A simple apparatus places the ends of a capillary in two buffer reservoirs containing electrodes. An applied electric potential difference drives migration along the tube; a detection window interrogates the passing material. In the uniform-field model, with total capillary length , detector distance and signed applied potential ,
The last expression requires net motion towards the detector. Thus a neutral marker can move with electroosmotic flow even though its electrophoretic mobility is zero. The instrument's distinction between total length and detector distance matters when converting migration times into apparent mobility.[ref]
For an original ideal example, take , and . Two net mobilities give speeds and and migration times 250 and 200 s. These are model values, not assigned chemical identities. Band widths from diffusion, injection and other dispersion mechanisms determine whether the time difference is actually resolved.
Narrow tubes also help remove Joule heating. For a uniform cylindrical liquid conductor, heat generation per volume is , with electrical conductivity , while wall area per liquid volume is for tube radius . This geometric scaling motivates small bores, although temperature control still depends on heat transfer through the wall and surroundings. Increasing the electric field indefinitely is therefore not a free route to perfect separation.
A primary 1981 demonstration used a 1 m glass capillary with 75 micrometre internal diameter, a 30 kV applied potential and a fluorescence detector. It resolved fluorescent derivatives of amino acids, peptides and urinary amines. The observed result was separated fluorescent zones; the need to derivatize the samples was an explicit detection limitation.[ref]