Electron multiplier Created 2026-10-18 Updated 2026-11-05
An electron multiplier amplifies a signal by accelerating incoming electrons into surfaces that emit additional secondary electrons. Repeating this process at successive stages produces a measurable charge packet from a small initial signal.
A photomultiplier adds a photosensitive input that converts incident light into a photoelectron; its dynodes then perform multiplication. In a simple equal-stage model with mean secondary yield , stages give mean gain . The output fluctuates because emission is stochastic, and saturation can invalidate constant gain. Multiplication increases the signal scale but cannot recreate input events that were never detected.
Mass spectrometer detector 2026-11-05
A mass spectrometer detector turns arriving ions into a recorded electrical signal. It is distinct from the mass analyzer that makes mass-to-charge ratio information distinguishable. The measured response can depend on collection, conversion and amplification as well as on how many ions arrived.
A Faraday cup collects electric charge directly. Ideally, a flux of ions with known charge gives . For singly charged positive ions per second, the current is , about . Escaping secondary electrons and background currents must be accounted for when inferring particle delivery.[ref]
An electron multiplier instead uses secondary emission to amplify the initial conversion signal. The manufacturer's detector description shows discrete dynodes, a continuous-dynode alternative, and a conversion dynode before the multiplier.[ref] In an illustrative equal-stage model, ten stages with mean yield give mean gain . This is a model calculation, not a measured gain for a named instrument.
Detector gain and detection efficiency answer different questions: the size of an output event and the probability of registering an arriving ion. Increasing gain cannot recover missed ions. At high signals, mass spectrometer detector saturation can also invalidate proportionality between ion delivery and recorded response. Quantitative peak comparisons therefore need a tested response range and appropriate instrument calibration, not merely a large signal.
Paul trap Created 2026-10-18 Updated 2026-10-20
The Paul trap alternates focusing and defocusing electric forces. Their time average along a moving trajectory need not vanish, even when the electric field averages to zero at every fixed point. A spatial gradient makes the fast displacement correlate with the driving electric field; the RF-averaged trapping potential calculates the resulting restoring force.
Wolfgang Paul's lecture describes hyperbolic ring/endcap electrodes, loading by electron-impact ionization of low-pressure gas, and resonant ejection through an endcap hole into an electron multiplier. Sweeping RF voltage produces an air mass spectrum.[ref]
Electromagnetic traps for charged and neutral particles by Wolfgang Paul. Paul, pp. 608–614, Figs. 6, 11; Columbia scan