Magnetic resonance velocity encoding makes a component of motion measurable through the phase of a nuclear magnetic resonance signal. Controlled magnetic field gradients give different phase shifts to spins with different trajectories. Phase-contrast magnetic resonance imaging compares encodings to extract a velocity component, rather than interpreting ordinary image brightness as speed.[ref]
In an original idealized model, assume constant velocity along one encoding axis and effective gradients with equal gradient zeroth moments. Then the encoding difference obeys
where is the angular gyromagnetic ratio and is the gradient first moment difference. For , and , this gives . These stipulated values illustrate the conversion, rather than a particular scanner experiment.
Reversing the moment difference reverses the phase sign under this convention. Reducing its magnitude decreases sensitivity to a given velocity while increasing the velocity encoding limit. A measured voxel can contain a range of velocities, and background phase or phase aliasing can complicate reconstruction.
The encoded component also differs from volumetric flow rate: estimating transported volume requires combining the appropriate normal velocity with area and time, not merely recording one peak speed.