Mass spectrometry imaging associates mass spectra with positions in a specimen, allowing an image to be formed for a chosen mass-to-charge ratio or spectral feature. Instead of reducing the whole specimen to one spectrum, it retains a link between molecular signal and location.
A 1997 experiment used matrix-assisted laser desorption/ionization, MALDI, to collect an ordered array of spectra from coated tissue sections or tissue imprints. Selected-ion images mapped peptides and proteins; one demonstration localized insulin in rat pancreatic tissue.[ref]
Molecular imaging of biological samples: localization of peptides and proteins using MALDI-TOF MS by R. M. Caprioli and others. Abstract, sample coating, ordered spectra and tissue demonstrations
The apparatus couples controlled sampling positions with a mass spectrometer; an image pixel records a spectral measurement, not a conventional colour photograph.An original two-position example shows why interpretation depends on scaling. A selected feature gives raw signals 10 and 20, while the respective summed spectral signals are 100 and 400. The second position is twice as bright in the raw feature image. Dividing each feature by its local total gives and , reversing which position has the larger fraction. Mass spectrometry image normalization changes the question being displayed; neither version alone is a universal concentration map.
Spatial sampling, material redistribution during preparation, ion production and detector response can all affect the result. A common peak coordinate does not by itself resolve every molecule with the same nominal mass or establish biological function. Retaining spectra, coordinates and processing choices makes the image scientifically interpretable.
Mass spectrometry image normalization rescales the spectra associated with individual image pixels to make a specified comparison more meaningful. In mass spectrometry imaging, raw intensity depends on both the sample and measurement conditions; MALDI matrix distribution can change the recorded intensity. A normalization rule therefore embodies an assumption about which variation should be removed.[ref]
Normalization in MALDI-TOF imaging datasets of proteins: practical considerations by Sören-Oliver Deininger and others. Introduction and “A pledge for normalization”: preparation-dependent signal variation and the assumptions of spectrum-wide scaling.
For example, total ion current normalization divides each nonzero spectrum by the sum of its included intensities. Write pixel as values and setIn an invented two-feature example, spectra and both become . This removes a common multiplicative change. But becomes : the first feature now appears half as strong as in even though its raw intensity is unchanged. The normalized image represents a relative share of the included signal. It answers a different question from an image of raw intensity or calibrated molecular abundance.
In actual MALDI datasets, Deininger and colleagues found that locally dominant signals could introduce artifacts under total-signal normalization; they compared median and noise-based alternatives. Their examples motivate inspecting the denominator as well as the selected ion image.[ref]
Normalization in MALDI-TOF imaging datasets of proteins: practical considerations by Sören-Oliver Deininger and others. Abstract and Figures 2–4: differing tissue profiles and normalization-induced artifacts.
Specify the mass range, excluded peaks and zero-signal handling. Neither total ion current normalization nor another single scalar correction can, in general, undo selective suppression of one molecular signal.Total ion current normalization rescales a mass spectrum using its total included ion signal. For nonnegative, equally spaced spectral samples , a common implementation computes and reports when . The normalized values then sum to one. Multiplication by a common reference total gives another convention with the same relative ratios. The name refers to the summed spectral signal; a stored intensity need not itself be expressed in amperes.
This gives exact invariance to a common positive multiplier: replacing every by leaves every unchanged. It does not establish that every experimental disturbance acts through such a multiplier. The mass-to-charge ratio range, excluded peaks and treatment of background determine which signal contributes to the total ion current used for normalization.
A worked example exposes the effect of an unrelated dominant peak. Start with three recorded intensities , giving normalized values . If only the third intensity doubles, the spectrum becomes and the first normalized value falls to . The first feature's raw signal has not changed. Its normalized value has decreased because the denominator now contains more of a different feature. A spectrum of all zeros has no defined unit-sum normalization; assigning it an arbitrary denominator conceals missing information.
In MALDI tissue images, locally abundant peaks produced this kind of misleading attenuation, motivating comparisons with median- and noise-based normalization.[ref]
Normalization in MALDI-TOF imaging datasets of proteins: practical considerations by Sören-Oliver Deininger and others. “A pledge for normalization”, mouse-pancreas results and Figure 4: dominant insulin peaks and apparent attenuation of other signals.
Total ion current normalization can remove common intensity scaling, but ion suppression affecting only selected compounds violates that simple model. Report both the normalization rule and what the normalized image is intended to compare. Articles by others on the same topic
There are currently no matching articles.