Operating-point time constants with nonlinear airway resistance describe small deviations about a specified flow and volume trajectory. The finite ratio of pressure to flow does not generally give the resistance governing those deviations. Instead, local response uses the derivative of the pressure-flow curve at the operating flow.
In an original one-compartment model, let elastic lung compliance be locally constant and let resistive pressure be . With volume measured relative to an elastic reference,A baseline trajectory has flow . Linearizing for small perturbations givesWhen the local coefficients are constant, the time constant is . It is the differential airway resistance times compliance, rather than the secant ratio .
For the stipulated curve , with numerical pressure in centimetres of water and flow in litres per second, while . With , the local time constant is 1.2 s, not 0.8 s. A small added pressure step has volume responseAfter 0.8 s it has reached about 48.7% of its final increment; using the secant-based time constant would incorrectly predict 63.2%. The numbers describe the local model, not measured airway coefficients.
Baseline-trajectory dependence of local respiratory time constants is essential. An isolated filling compartment cannot maintain nonzero flow at a truly stationary volume. The example therefore refers to a driven baseline trajectory, or another arrangement with balancing flow. For an undriven stationary compartment, the baseline flow is zero. A large transient that substantially changes also changes , so one constant time scale need not describe its whole course.
A primary ventilator study compared mechanics measurements across inspiratory and expiratory methods and varied flow-related settings.[ref]
Static and Dynamic Measurements of Compliance and Driving Pressure: A Pilot Study by Pierre Tawfik and others. Introduction, methods C and D; Study Procedures
Report the operating state, perturbation size and pressure references with any inferred time constant. A single finite pressure-flow ratio does not establish a universal relaxation property.Baseline-trajectory dependence of local respiratory time constants means that the linearized response can change as volumetric flow rate and elastic state change during a breath. A time constant measured at one operating point need not describe the entire trajectory, even when the underlying constitutive laws stay fixed.
In an original one-compartment model, let pressure satisfy . Around a specified baseline , a small perturbation obeysThe frozen-coefficient local time constant is when both coefficients are positive. For zero perturbing drive, the actual response is insteadThis follows by an integrating factor for the linear differential equation. The accumulated relaxation rate, rather than one selected endpoint value, controls the decay.
Choose a stipulated driven baseline volumetric flow rate L/s with s. Let elastic compliance be 0.20 L per cm water and resistive pressure be in the corresponding numerical units. Then and . The local time constant decreases from 1.2 s toward 0.4 s as the baseline volumetric flow rate falls.
Writing numerical time , the homogeneous perturbation has the exact relative amplitudeAt one second it is about 0.322. Freezing the initial time constant would instead give ; freezing the limiting value would give . These are distinct approximations to a changing-coefficient problem, not measured patient responses.
The baseline here is explicitly driven: its external pressure must sustain the chosen trajectory. It is not claimed to be the natural unforced expiration law of the nonlinear compartment. The primary respiratory pilot study compares methods that depend on volumetric flow rate, pressure and expiratory time scales, and discusses why expiration need not have one exponential time constant.[ref]