Generic image registration estimates a coordinate transformation aligning corresponding features in two images. A rigid transform permits translation and rotation, an affine transform additionally permits linear deformation, and a nonrigid model permits spatially varying displacement. The transformation class expresses assumptions about what changes between the acquisitions.
One can minimize a discrepancy between and , using interpolated image values and an explicit optimization procedure. Squared intensity differences suit some comparable acquisitions; mutual information can compare statistical dependence when image contrasts differ. Landmarks, feature descriptors or additional geometric constraints provide other information. Smoothness or other regularization controls overly flexible transformations. A favorable objective value does not by itself prove correct feature correspondence.
TODO: Biology deferred at the user's request: biological radiation effects and clinical risk assessment.
In pulse-echo ultrasound, a transducer emits a pulse and records returning echoes. For a straight round-trip path in a medium of assumed sound speed , an echo delay gives depth . Reflections depend on acoustic-impedance changes, while attenuation, scattering and propagation geometry affect the received amplitude.
An A-mode ultrasound scan displays echo amplitude against inferred depth along one beam. A B-mode ultrasound scan maps received amplitude to brightness and assembles many beam directions or translated beam lines into a spatial image. An M-mode ultrasound scan repeats one beam line and displays its depth profile versus time, making motion along that line visible. Thus the display axes and acquisition pattern differ even when the pulse-echo principle is shared.
For a pulse lasting , a simple axial-separation estimate is . Lateral resolution depends on aperture, wavelength, focusing and depth. Shorter pulses and broader bandwidth can improve axial resolution; an assumed sound speed or a multiple-reflection path can instead misplace an echo. Brightness is not a direct universal map of composition.
Doppler ultrasound uses frequency or pulse-to-pulse phase changes from moving scatterers. In a simple backscatter geometry, . Inferring speed requires the beam-to-motion angle; pulsed sampling can alias frequencies, and an unknown angle leaves a velocity-projection ambiguity. This describes generic acoustic acquisition rather than a clinical diagnosis.
TODO: Biology deferred at the user's request: biological tracer behavior and clinical PET applications.
The generic physics of positron emission tomography uses electron-positron annihilation. A slowly moving electron-positron pair predominantly produces two nearly opposite photons with energy close to each. A ring of detectors records approximately simultaneous interactions; a coincidence identifies a coincidence line of response along which an annihilation could have occurred. Finite Positron travel before annihilation and nonzero pair momentum limit how directly this line identifies the original emission location.
Many such measurements constrain an emission distribution through a tomography model. Attenuation, scattered coincidences, unrelated random coincidences, finite detector resolution and dead time alter the observed counts and require calibration or modeling. With useful time-of-flight information, the arrival-time difference further localizes position along the line: a timing uncertainty corresponds to a position scale . The coincidence data therefore provide spatial constraints, rather than each event directly supplying a complete three-dimensional image point.