Photoresistor spectral calibration determines how a photoresistor's electrical reading depends on the illumination spectrum under specified geometry, background, temperature, bias and adaptation history. A resistance-versus-lux curve measured with one lamp is not automatically valid for another spectrum. The manufacturer's primary CdS datasheet specifies a source for its light-resistance tests and preconditions cells with light before electrical measurements.[ref]
NSL-5162 CdS photoresistor datasheet by Advanced Photonix. page 2, light-resistance test conditions and adaptation footnote
The distinction is between irradiance and illuminance. With spectral irradiance , illuminance is approximatelywhere is the photopic luminous efficiency function. The detector weights wavelengths according to its own physical response, which need not be proportional to . NIST's detector-based photometric realization measures that spectral response, rather than assuming an arbitrary detector reproduces the photometric weighting.[ref]
Realization of the candela. Detector-based realization, illuminance responsivity and spectral integrals
For a locally linear small-signal model about a fixed operating state, writeHere is the differential spectral responsivity of electrical conductance, including the stated illumination geometry. It is not automatically a global, illumination-independent material constant. A background-dependent photoresistor spectral response requires rechecking this derivative when the operating point changes.
An original two-band example makes the mismatch quantitative. Stipulate relative photometric weights , and equal detector generation weights for two narrow bands. Equal illuminance requires four times as much irradiance in band , so its detector-weighted generation is four times greater. If, over this illustrative range, , the resistance ratio is . These weights and exponent are stipulated, not measured specifications or tabulated values for named wavelengths. Equal lux can produce substantially unequal resistance.
A proposed spectral characterization uses narrow-band sources or a monochromator, a reference detector to measure incident irradiance, fixed beam geometry and controlled detector temperature. Record resistance or conductance after a defined settling time at several background levels. A separate calibrated photometer can establish illuminance; it cannot by itself establish the photoresistor's spectral weighting. Keep electrical loading and self-heating fixed while changing the source.
For a broadband transfer, characterize the reference and test spectra and test whether the measured spectral-response model predicts their relative readings. Existing spectral mismatch correction methods address the analogous transfer problem for photometers, but a nonlinear photoresistor first needs a justified operating-point model.[ref]
NIST Measurement Services: Photometric Calibrations by Yuqin Zong and others. §3.4.3, equation 22, and §3.6.2, spectral mismatch for sources different from the calibration lamp
The result is a calibration valid for stated conditions, not a universal conversion from resistance to optical power.A background-dependent photoresistor spectral response is the differential change of a photoresistor's output with wavelength-specific illumination, evaluated at a stated background state. It can depend on illumination level, spectral composition, bias, temperature and prior adaptation. The primary CdS datasheet specifies light-adaptation preconditioning before electrical tests, demonstrating that a measurement condition includes more than the instantaneous lux value.[ref]
NSL-5162 CdS photoresistor datasheet by Advanced Photonix. page 2, light-adaptation footnote and test-source conditions
Let the steady electrical conductance be a functional of incident spectral irradiance. About a background , define the local response byThis derivative concerns small changes after specifying the operating state. A finite change or rapid modulation needs further nonlinear or dynamic characterization. Photoresistor spectral calibration establishes the required optical and electrical conditions.
An original separable model makes a useful distinction. Suppose and , with positive . ThenThe amplitude changes with background, but the normalized spectral shape remains . With , quadrupling multiplies the absolute differential response by . For a small one-percent generation increment, the leading fractional conductance change above the dark contribution is ; the exact model gives .
A change in spectral shape requires more than that single separable nonlinearity. For an illustrative two-band response in dimensionless input coordinates, letThe derivatives are and . Changing the first-band background from to changes their ratio from to . This toy model shows how relative spectral sensitivity can change; it is not measured behavior of a named component. Separable and nonseparable photoresistor spectral-response models should consequently be distinguished in a calibration analysis.
Resistance readout adds another operating-point factor: since , . A resistance-divider output has yet another derivative. Comparing raw electrical slopes without converting to the same output variable can imitate a change in material response. Likewise, equal illuminance need not mean equal detector excitation, because the photopic luminous efficiency function supplies a different spectral weighting.[ref]
Realization of the candela. Detector-based realization, illuminance responsivity and spectral integrals
A proposed apparatus combines a stable background source with a weak wavelength-selectable probe beam, a monitored reference detector and a fixed illumination geometry. At each background, measure the output derivative while decreasing probe amplitude to test the small-signal regime, then repeat after the prescribed settling or adaptation period. Normalize spectra only after retaining their absolute scale. This separates a changing overall gain from a changing spectral shape and from readout loading; it does not assume that a curve measured at one background transfers to every illumination condition.