Burette Created 2026-10-14 Updated 2026-10-18
A burette delivers an adjustable measured volume of a chemical solution; the change in its reading gives the delivered volume in an acid-base titration.
Its downward-increasing scale measures a change in reading, not the initial liquid inventory. For initial reading and final reading without refilling, the nominal delivered volume is . An illustrative to reading therefore represents . There is no requirement to start at zero. An air bubble disappearing from the tip uses part of the apparent delivery to fill the tip, and a hanging drop has not yet entered the sample; both break the intended volume account.
Figure 1.
A pipette selects the sample and a burette measures added titrant by two liquid meniscus readings
. Original schematic, not a photograph or experimental readout. The apparatus shows a clamped burette, mixed sample, and optional glass electrode beside the stirring bar. Enlarged illustrative scale readings increase downward; the bottom of the curved liquid meniscus gives and . Their difference is the added volume. Details.
For transparent aqueous samples, read the bottom of the liquid meniscus at eye level; viewing from above or below shifts its apparent alignment. The mark is a comparison between a curved liquid surface and a scale, not a direct measure of amount of substance. Eye-level reading and the distinction between contained and delivered volumes are part of solution preparation and volumetric calibration.[ref]
Both readings contribute measurement uncertainty. If each has independent standard deviation , their difference has standard deviation ; a common additive reading offset cancels, while a scale-factor error remains. For the inventory , uncertainty in titrant concentration remains even with perfect readings. This is why repeat delivery, independent volume calibration, and titrant standardization answer different questions.
Titrant standardization 2026-10-18
Titrant standardization determines the actual titrant concentration by reaction with a known inventory, often a primary standard. It is distinct from calibrating the delivered volume or a glass electrode's response. For a known monoprotic acid inventory and a suitable blank-corrected endpoint,
An original illustrative standard of requiring , with matched blank , gives . The measured quantities here are stipulated examples. Independent solution preparation and volumetric calibration tests the pipette, burette, and volumetric flask; titrant standardization tests the reacting inventory per delivered volume.
The useful chain is known mass, known composition, known amount of substance, and then a measured titre through a specified stoichiometry. If each standard unit consumes units of titrant, the general inventory equation is , where is the corrected titre. The one-to-one formula above has . A titration endpoint supplies an observable stopping rule; whether it represents the equivalence point accurately remains a separate question. Standardization endpoint and blank bias makes the direction of the resulting error explicit.