Steve R. Domen developed physical water calorimetry methods at the National Institute of Standards and Technology. His 1994 sealed-water instrument measured absorbed dose from a minute radiation-induced temperature rise, while investigating how dissolved gases change the fraction of deposited energy that appears as heat.[ref]
A Sealed Water Calorimeter for Measuring Absorbed Dose by Steve R. Domen. Title-page byline, abstract and section 1
The apparatus enclosed water in a thin glass vessel immersed in a larger water tank. Two calibrated thermistors in glass capillaries formed part of a Wheatstone bridge that detected the temperature response. The vessel also limited convection. Comparing different dissolved-gas conditions tested the heat defect, rather than assuming that all absorbed energy immediately becomes heat.[ref]
A Sealed Water Calorimeter for Measuring Absorbed Dose by Steve R. Domen. Sections 2.1–2.3, figures 1–4; abstract, approximately 3700 measurements
For an original ideal calculation, neglect the heat defect and heat transfer during a reading. With water specific heat capacity and measured rise ,This illustrates why sub-millikelvin thermometry matters. It is not a reported run: the real measurement model must include the instrument's corrections and measurement uncertainty.