The sealed-water calorimeter of Steve R. Domen is an apparatus for measuring absorbed dose through the radiation-induced temperature rise of a controlled water volume. Domen's 1994 design enclosed the inner water in a thin-walled cylindrical glass vessel about 110 mm long and 33 mm in diameter, immersed in a larger water bath. The vessel also acted as a barrier against convection.[ref]
A Sealed Water Calorimeter for Measuring Absorbed Dose by Steve R. Domen. Section2.1, Design: vessel dimensions, surrounding water and convective-barrier role.
Two calibrated thermistors served as temperature sensors in opposite arms of a Wheatstone bridge, increasing the electrical response to their common temperature rise. The apparatus therefore connected a small thermal change to a measurable resistance change, rather than directly counting radiation events.[ref]
A Sealed Water Calorimeter for Measuring Absorbed Dose by Steve R. Domen. Introduction, Eq1 and Section2: two-thermistor bridge and temperature conversion.
A central experiment compared water fillings with different dissolved-gas conditions. The temperature response could change because radiation also drove chemical reactions, motivating investigation of the heat defect. This comparison illustrates a useful apparatus principle: controlling the geometry and reading the thermometer carefully does not, by itself, establish that every joule deposited in the water becomes measurable heat.[ref]
A Sealed Water Calorimeter for Measuring Absorbed Dose by Steve R. Domen. Abstract and results: repeated measurements with different fillings and heat-defect investigation.