A stimulated echo is a delayed nuclear magnetic resonance signal produced by a sequence that stores part of a dephased magnetization pattern along the static magnetic field and later converts it back into detectable transverse magnetization. In a common ideal sequence, three pulses are applied at times , and . The selected stimulated echo appears a further after the third pulse, at .[ref]
The first interval lets spins with different Larmor frequencies accumulate different phases. The second pulse converts part of that frequency-dependent pattern into longitudinal magnetization. During the storage interval this component does not precess as transverse magnetization does. The third pulse returns part of the pattern to the transverse plane, where the appropriate contribution rephases. Storage retains a frequency-dependent population pattern, not a continuously rotating transverse vector.[ref]
For an original ideal calculation, hold fixed and neglect diffusion, pulse errors and exchange. If is the spin-lattice relaxation time, the selected echo contribution obeys
A storage interval therefore halves that contribution. The two transverse intervals still incur transverse relaxation; longitudinal storage does not undo every source of signal loss.
This distinction is useful for diffusion measurements when a longer observation interval is wanted without keeping the relevant pattern transverse throughout. A 1989 experiment used pulsed-gradient stimulated echo sequences to reduce particular background-gradient effects in heterogeneous samples. It tested specified pulse arrangements, rather than establishing that every stimulated echo removes every gradient error.[ref]