Pulsed field gradient NMR, or PFG NMR, uses timed magnetic field gradients to encode position in the phase of a nuclear magnetic resonance signal. An appropriately refocused pair of encoding episodes cancels the position-dependent phase for stationary spins. Motion between the episodes can prevent cancellation, making the echo sensitive to displacement.
For an original narrow-pulse derivation in one spatial direction, let each effective encoding pulse have gradient magnitude and short duration , and write , where is the angular gyromagnetic ratio. Neglect displacement during each pulse. For positions separated by an observation interval , a choice of encoding sign gives . The normalized signal is the ensemble average of .
For free diffusion with diffusion coefficient , displacement has a zero-mean normal distribution of variance . Its characteristic function givesFor invented values , and , the attenuation is . Finite pulse duration, restricted motion and other sequence details require their corresponding model rather than automatic use of this narrow-pulse result.
The 1965 experiment of E. O. Stejskal and J. E. Tanner combined a derivation for time-dependent gradients with experimental tests. Reducing gradients during radio-frequency pulses and echo detection relaxed practical limitations of a continuously strong gradient; the authors measured slow diffusion in dry glycerol.[ref]
Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field Gradient by E. O. Stejskal and J. E. Tanner. Page288, Abstract; Theory and experimental tests
PFG NMR can also use a stimulated echo for storage between encoding periods. Distinguishing motion-induced attenuation from other signal losses is part of the measurement, not a consequence of the method's name alone.
The diffusion weighting b-value quantifies the encoding applied by a magnetic field gradient waveform in a diffusion measurement. For a refocused experiment in one spatial direction, defineHere is the angular gyromagnetic ratio, in radians per second per tesla, and includes the sign reversal caused by the refocusing pulse. The echo time is , with for refocusing of stationary spins. Thus has units and has units . The squared encoding history matters even though its final value is zero.[ref]
Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field Gradient by E. O. Stejskal and J. E. Tanner. Pages289–290: time-dependent-gradient theory and refocusing
For unrestricted, isotropic Gaussian diffusion with constant diffusion coefficient , the otherwise matched signal ratio is . Here uses the same pulse timing and detection conditions with the applied diffusion-encoding gradients removed. The diffusion weighting b-value describes the experiment's weighting; describes the assumed sample model.
For an original unit check, . With , the ideal attenuation is . Quoting a numerical without its units can therefore cause a millionfold scaling error.
The Stejskal-Tanner equation gives for its equal rectangular pulses. More general waveforms require their own integral. For direction-dependent diffusion, a scalar alone cannot specify the complete directional encoding and response; a tensor description is needed.
The Stejskal-Tanner equation relates diffusion-sensitive echo attenuation to a rectangular gradient-pair experiment:It assumes free Gaussian diffusion with a constant scalar diffusion coefficient , ideal refocusing and two equal nonoverlapping rectangular gradient pulses of magnitude and duration , whose starting times are separated by . The gyromagnetic ratio is angular. The baseline is an otherwise matched echo without the applied diffusion-encoding gradients. These definitions are part of the formula, rather than optional choices of notation.[ref]
Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field Gradient by E. O. Stejskal and J. E. Tanner. Pages289–290: Theory and rectangular-gradient case
The finite-duration term follows directly from the encoding history. Write . The effective encoding rises linearly from zero to , stays constant for , then falls linearly to zero. Each ramp contributes to the diffusion weighting b-value, soThe two physical gradients can have the same polarity: the intervening pulse reverses the effective phase-encoding sign.
For the illustrated and , the normalized integral is , rather than . This isolates the pulse-duration correction without assigning an unmeasured gradient amplitude.
The 1965 work combined theory with apparatus tests, including slow self-diffusion in dry glycerol.[ref]
Spin Diffusion Measurements: Spin Echoes in the Presence of a Time-Dependent Field Gradient by E. O. Stejskal and J. E. Tanner. Pages288 and291–292: motivation, apparatus and experiments
Restricted diffusion, flow, background gradients and measurement noise require their own interpretation; fitting a single attenuation to this equation does not establish all of its assumptions. Articles by others on the same topic