Coincidence line of response Created 2026-10-20 Updated 2026-10-25
A line defined by an approximately coincident detector pair in annihilation-photon tomography. An event constrains a possible annihilation location to that line within instrumental and physical uncertainties. Time-of-flight data can add a position constraint along it; attenuation, scattering and random coincidences change the forward model.
Inverse problem Created 2026-09-24 Updated 2026-10-14
An inverse problem reconstructs an unknown object from observed data and a specified model . The associated forward problem predicts the observations from . In image deblurring, contains sharp pixel values and a matrix mixes neighbouring pixels; in tomography, the observations are integrals through the object. The appeal is that indirect measurements can reveal otherwise inaccessible structure. The difficulty is that the forward problem may erase precisely the distinctions we want to recover.
Consider this completely computable two-pixel image deblurring model:Each measured pixel retains of its own value and receives of its neighbour's. Adding measurement noise gives , whose direct reconstruction is . The Euclidean norm of the reconstruction error is , ten times the data error . A negative second pixel appears even though both observed pixels are positive.
Despite this wrong reconstruction, its residual of an inverse problem is exactly zero: the predicted data equal the noisy observations. Thus matching the data perfectly does not certify reconstruction accuracy. The relevant question is how large a change in the unknown could hide within the measurement uncertainty. A condition number quantifies that sensitivity for an invertible matrix; the null space identifies changes that no improvement in precision can reveal.
The exact singular value decomposition explains this amplification. With the orthonormal basisthe mean mode survives unchanged while the difference mode shrinks tenfold. Direct inversion multiplies difference-mode measurement noise by ten. The condition number is . The beauty of the singular value decomposition is this separation into independent channels: it identifies which information survives and which needs protection by regularization of an inverse problem.
Past exam of the medical physics course of the University of Cambridge 2012 Paper 1 3 d Solution 2026-10-20
TODO: Biology deferred at the user's request: biological tracer behavior and clinical PET applications.
The generic physics of positron emission tomography uses electron-positron annihilation. A slowly moving electron-positron pair predominantly produces two nearly opposite photons with energy close to each. A ring of detectors records approximately simultaneous interactions; a coincidence identifies a coincidence line of response along which an annihilation could have occurred. Finite Positron travel before annihilation and nonzero pair momentum limit how directly this line identifies the original emission location.
Many such measurements constrain an emission distribution through a tomography model. Attenuation, scattered coincidences, unrelated random coincidences, finite detector resolution and dead time alter the observed counts and require calibration or modeling. With useful time-of-flight information, the arrival-time difference further localizes position along the line: a timing uncertainty corresponds to a position scale . The coincidence data therefore provide spatial constraints, rather than each event directly supplying a complete three-dimensional image point.
Positron emission tomography 2026-10-20
Tomography using approximately coincident annihilation-photon detections to constrain an internal emission distribution. A calibrated model accounts for propagation, detector response, unrelated coincidences and scattering.
Single-photon emission computed tomography 2026-10-14
A form of tomography that reconstructs an internal radioactive source from directional counts of individually emitted photons. Known spatial photon absorption gives the exponential path weight in an attenuated Radon transform. Full directed data and a suitable reconstruction formula recover the source intensity rather than a transmitted exterior-beam attenuation map.