A water phantom is a water-filled reference object used to measure how a radiation field deposits energy under specified, reproducible conditions. In radiation dosimetry, a detector can be positioned at controlled depths and lateral locations inside the water. The measured quantity is related to absorbed dose in that reference medium; the tank is not a complete anatomical model of a patient.[ref]
Absorbed Dose Determination in External Beam Radiotherapy by the International Atomic Energy Agency (2000). Section4.2.3, Phantoms, printedpp.39–40
The surrounding medium matters physically. Photons and charged particles can scatter, attenuate and transfer energy before it is collected locally. An empty holder, a small isolated water sample and a large water tank need not produce identical fields at a detector. The named IAEA edition treats both water as reference material and the margins of scattering material around a measurement region. A water-equivalent material must be compared in relevant absorption and scattering properties, not identified by similar bulk density alone.[ref]
A practical apparatus combines a tank, a positionable ionization chamber or another characterized detector, a readout and a way to define the water surface. For example, PTW's MP1 description specifies one-dimensional detector travel, levelling supports and fill marks. Its depth-positioning mechanism supports a depth-dose curve; one mechanical coordinate is not itself a full three-dimensional scan.[ref]
MP1 water phantom by PTW. General Information and Key Features
An original position-error model shows why the reference coordinates matter. At a fixed depth, stipulate a lateral absorbed dose profileA displacement from changes the dose ratio toThe same displacement from the centre gives . Thus the same absolute positioning error can produce about variation on a gradient and only at the centre of this stipulated profile. These are mathematical examples, not measured beam characteristics or equipment accuracy claims.
An ionization chamber also samples a finite sensitive volume, so detector volume averaging can differ from a point value of . Converting its collected electric charge to absorbed dose requires the applicable measurement calibration and measurement model, as described in medical physicist. Changing the tank position does not remove a detector's intrinsic spatial response.
Comparisons should therefore identify the water surface, detector reference point, irradiation geometry, water conditions and background treatment. The IAEA discussion of waterproof sleeves and chamber positioning illustrates why the detector and its surrounding hardware are part of the measurement configuration.[ref]