FRET sensitized emission is FRET acceptor fluorescence caused by transfer from an excited FRET donor. Excite predominantly the FRET donor and collect both emission bands. Reduced FRET donor emission and increased FRET acceptor emission are useful together, but an acceptor-to-donor emission ratio is not automatically the FRET efficiency: brightness, label abundance and channel collection also enter.[ref]
Advanced Fluorescence Microscopy Techniques—FRAP, FLIP, FLAP, FRET and FLIM by Hellen C. Ishikawa-Ankerhold and others. §3.4.2: sensitized emission and dependence on optical correction and label concentrations
FRET optical crosstalk includes FRET donor emission leaking into the FRET acceptor detection band and direct excitation of the FRET acceptor by the nominal FRET donor excitation light. Measure donor-only, acceptor-only and unlabeled preparations with the same fluorescence microscope settings to separate these contributions and background.[ref]
Advanced Fluorescence Microscopy Techniques—FRAP, FLIP, FLAP, FRET and FLIM by Hellen C. Ishikawa-Ankerhold and others. §3.4.2, samples i–iii and equations 20–21: spectral correction controls
An explicit linear measurement model makes the controls' purpose clear. Let be background-subtracted donor-band counts under donor excitation, acceptor-band counts under acceptor excitation, and acceptor-band counts under donor excitation. Donor-only and acceptor-only preparations determine leakage coefficients and . Thenis a corrected FRET sensitized emission signal under the assumed linear, fixed optical response. In a synthetic example , , , and give . Even a nontransferring sample would give from these two contaminants. Correcting that signal still does not convert into FRET efficiency without a calibration of excitation, yields and detection.
A biological comparison also needs a pairing-negative preparation and a known transferring positive preparation. Match label abundance and localization where possible; changing concentration can change encounter frequency as well as brightness. Check that attaching the labels preserves the protein function being investigated. Otherwise the measurement can faithfully report a perturbation introduced by the labels themselves.