In scanning tunneling microscopy, under simplified constant-tip-density assumptions, with . A constant-current height map mixes actual surface height with local electronic density, barrier/work-function changes and tip structure. Changing bias can change contrast without moving atoms. Scanning tunneling spectroscopy instead records energy dependence, often through ; tip states, matrix elements and thermal or modulation broadening still affect its interpretation.
In atomic force microscopy, contrast comes from tip-sample forces or their gradients: short-range chemical bonding and repulsion, dispersion forces, electrostatics and sometimes magnetism or capillary forces. Contact deflection, oscillation amplitude and frequency shift are different observables. A force-gradient image can therefore differ from a geometrical-height image, and changing the tip termination can strongly change atomic contrast.
STM's exponential distance sensitivity can give excellent vertical sensitivity; usable resolution also depends on current noise, feedback, vibration and drift. Lateral contrast depends on the apex orbitals and electronic spatial extent, not merely the nominal tip radius. In AFM, broad-tip geometrical convolution broadens larger features; a sharp terminal atom and short-range forces can produce much finer atomic contrast. Cantilever stiffness, amplitude, frequency noise, force range, thermal fluctuations and feedback influence height sensitivity. In both techniques, detecting a small displacement is distinct from resolving two neighboring objects.
An example is the Cu(111) surface-state standing-wave experiment. Low-temperature STM and spatially resolved spectroscopy connect atomic steps or defects in a surface map with surrounding energy-dependent electronic interference patterns. Their period changes with bias, whereas a fixed structural corrugation would not acquire that electronic dispersion. The result shows why spectroscopic maps reveal surface-electron scattering beyond what a height image alone identifies.[ref]
Imaging standing waves in a two-dimensional electron gas by M. F. Crommie and others. 1993, observation of Cu(111) local-density-of-states standing waves
The apparent fringes are not a second set of equally spaced surface atoms.Transmission electron microscopy accelerates electrons through a thin sample, then uses electron lenses to form an image or diffraction pattern. Elastic scattering encodes projected structure and potential; inelastic scattering carries excitation and chemical information. Thin specimens reduce unwanted multiple scattering and absorption, although coherent dynamical scattering can still matter in a crystal. High-resolution phase-contrast images depend on defocus and the transfer function as well as the specimen; they are not direct height maps.
The relativistic wavelength is . A shorter wavelength permits fine resolution but does not guarantee it. A schematic probe/image blur budget contains diffraction , spherical aberration and chromatic aberration , plus source size, instabilities, specimen motion and radiation damage. Optimizing the aperture angle balances these terms; increasing accelerating voltage can also alter damage and contrast.
Bright-field imaging accepts the central transmitted beam or its neighborhood. Regions that scatter electrons outside that acceptance can appear dark. Conventional dark-field imaging selects a diffracted beam. Annular dark-field electron imaging in a scanning TEM instead collects a range of scattered angles while a probe is scanned. A sufficiently high-angle detector often gives strong atomic-number sensitivity and more nearly incoherent contrast; thickness, channeling and detector geometry prevent a universal conversion of intensity to composition.
Electron energy loss spectroscopy measures energy lost by transmitted electrons. Core-ionization thresholds identify elements, while near-edge fine structure probes bonding and unoccupied states. At successive probe positions, one can record a spectrum image or form a map from a selected edge after background subtraction. Plural scattering, edge overlap, delocalization, specimen thickness and counting statistics limit the interpretation. Chemical specificity comes from excitation energies and spectral structure, not from brightness alone.
A development already established before this 2012 paper is aberration-corrected transmission electron microscopy: multipole correctors compensate leading lens aberrations, allowing useful larger angular acceptance and finer probes. Correction does not remove every chromatic, stability or specimen limit. A concrete application was atom-by-atom annular-dark-field analysis of monolayer boron nitride with substitutional defects, resolving chemical differences between sites in a nonuniform lattice.[ref]
Atom-by-atom structural and chemical analysis by annular dark-field electron microscopy by O. L. Krivanek and others. 2010, low-voltage aberration-corrected STEM of monolayer boron nitride
Combining structural maps with spatially resolved core-loss spectra supplies another route to distinguish composition and bonding at interfaces. Both exploit a calibrated imaging or spectroscopy response rather than assuming every visible spot is chemically self-identifying. Articles by others on the same topic