In Fourier transform spectroscopy, a scanning Michelson interferometer measures intensity against optical path difference rather than selecting one narrow wavelength at a time. After subtraction of the constant background, an ideal stationary source gives an interferogram
A cosine Fourier transform recovers the spectral power density after instrumental calibration. A mirror displacement changes the round-trip path by ; the zero-path-difference burst aligns the scan, and a reference laser can calibrate the displacement.
Finite maximum path difference multiplies the interferogram by a window and convolves the recovered radiation spectrum with that window's Fourier transform. Thus is of order ; precise factors depend on scan range and resolution convention. Uniform path samples of separation need to avoid aliasing without bandpass information. interferogram apodization reduces sidelobes at the price of broader lines. The broad accepted optical throughput can be advantageous; the multiplex signal-to-noise advantage depends on detector noise and is not universal in photon-noise-limited measurements. Source drift during a scan, finite beam divergence, imperfect sampling and wave phase errors require control. The finite-window reasoning is the delay-range resolution in Fourier transform spectroscopy.[ref]
A finite optical pupil spreads a point into a point spread function. In incoherent imaging an extended intensity distribution is convolved with that function; its Fourier transform is the optical transfer function, which attenuates and ultimately removes high spatial frequencies. For an ideal circular pupil the point response is an Airy disk. The Rayleigh criterion for two equally bright incoherent points is
A microscope's object-space scale is of order , with the numerical factor depending on illumination and the criterion; measures the accepted ray cone. A larger pupil or shorter wavelength improves the diffraction scale. Aberrations, defocus, detector pixel sampling, motion, background and photon noise can worsen practical performance. Merely magnifying the same blurred image adds no information. Coherent sources require amplitude superposition instead of the incoherent intensity rule. Rayleigh resolution is a convention for separating a pair, distinct from the precision of localizing one isolated point; information does not abruptly vanish below it.
Boundary conditions quantize the transverse wavevector of an electromagnetic waveguide. For an ideal hollow rectangular conducting guide of sides , filled with a nondispersive medium of wave speed , a transverse electric mode of a rectangular waveguide or transverse magnetic mode of a rectangular waveguide obeys
The allowed integer pairs differ: TE permits one zero index, but not both; TM requires both positive. Here is the longitudinal propagation constant. Below cutoff is imaginary, so the mode is evanescent. Above cutoff,
The positive branch approaches the straight line far above cutoff, while near cutoff its slope tends to zero. Different frequencies acquire different group delays and a pulse can spread. A wave phase speed above does not convey superluminal information; in this ideal guide energy transport follows the group velocity. A single hollow conductor cannot support a TEM mode; two-conductor guides can have a zero-cutoff TEM branch. Material dispersion, dielectric fibres and lossy boundaries require their actual dispersion relation and need not preserve the product rule.