TODO: biomolecular force assays are user-deferred. The following notes cover the generic physical apparatus, calibrated dielectric handles and synthetic attachments only.
In optical tweezers, a tightly focused laser produces a restoring gradient force on a suitable transparent bead. Scattering also transfers momentum along propagation, so a stable three-dimensional trap needs axial as well as lateral restoration. An objective with high numerical aperture, correctly filled entrance pupil, controlled beam polarization and low-aberration sample chamber are practical design elements. Collect transmitted/scattered light with a condenser and measure bead displacement with calibrated imaging or back-focal-plane photodetection. A steerable beam or translation stage sets relative positions; the force signal comes from bead displacement, not commanded stage displacement alone.
Near a stable operating point, use a harmonic optical trap with . For a slowly changing external force, in that coordinate; the optical restoring force is . Calibrate each axis and the range over which its response is linear. Detector sensitivity must also be known: write , with measured detector voltage , offset and position sensitivity in . A spring constant in arbitrary detector units is not a force calibration.
Three complementary physical calibrations are useful. At thermal equilibrium the equipartition theorem gives
This needs an independently calibrated position scale and a variance corrected for drift, detector noise and exposure averaging. In an overdamped Newtonian medium, the optical trap Langevin calibration obeys
Using a one-sided position power spectral density per hertz,
The corner frequency and spectral amplitude test both relaxation and fluctuation size. If drag is known, . The isolated-sphere value must be corrected near boundaries; otherwise measured height dependence can be mistaken for a changed trap stiffness. An independent steady optical trap drag calibration moves fluid past the trapped sphere: . Use small positive and negative speeds to verify slope, offset and force range.
One can also determine position sensitivity without assuming the bulk drag. In a low-frequency sinusoidal-stage optical trap calibration, let the known local fluid velocity amplitude be at drive angular frequency . The coherent bead displacement amplitude is
If the corresponding detector amplitude is , then
After removing the coherent drive line from the thermal record, combine this sensitivity with equipartition theorem to recover , and then infer from . A published calibration combines thermal spectra and stage-driven motion without taking bead drag as an input.[ref]
Controls determine whether these formulas measure the apparatus rather than its noise. Record blocked-light offsets and independent fixed-bead noise; calibrate length with known motion; keep sampling and electronics bandwidth adequate around ; and include exposure filtering and aliasing in a spectral fit. motion blur reduces measured variance and therefore makes naive equipartition theorem overestimate stiffness; additive position noise has the opposite bias. Changing exposure time provides a direct diagnostic.[ref] Check temperature and laser heating, wall separation, bead size and refractive index, alignment and power stability. Repeat after changing power or optical geometry and compare thermal, spectral and drag estimates. For a synthetic force handle, also measure the attachment's compliance and geometry so that bead motion is not automatically identified with its extension.
Figure 1.
Optical trap calibration
. Details.