Spherical irradiance counts incident radiance from all directions without selecting a receiving-plane normal. Using the projected-area convention specified in the CIE entry's note2,Here is an element of solid angle. An ideal small sphere of radius intercepts radiant flux . The denominator is its diametrical projected area, not its full surface area.[ref]
International Lighting Vocabulary by CIE. Entry17-21-054, note2, small-sphere construction; note5, units
For an original isotropic radiance field example, take in every direction. Then . A one-sided plane instead receivesThe factor four follows from the different angular geometries; it is not a unit conversion. Averaging the sphere's intercepted power over its actual surface also gives , rather than the projected-area-normalized quantity above.
An all-direction receiver must have a characterized angular response. A single flat irradiance detector inherently samples a hemisphere with projected-area weighting, so rotating it or enclosing it in a diffuser does not supply an all-direction value without an appropriate response model and measurement calibration. Report the angular and area conventions with the result; equal units do not make irradiance and spherical irradiance interchangeable.