Radiometry measures the energy and power carried by electromagnetic radiation, including light. It asks how much radiation a source supplies, where it travels, and how much reaches a receiving surface. The measurements are physical rather than weighted by the sensitivity of the human eye: equal radiant fluxes at different wavelengths need not look equally bright. [ref].
The five quantities below answer different questions about the same radiation. The International System of Units makes the distinctions visible: a watt is a joule per second, while a steradian measures solid angle. [ref].
QuantityWhat is resolved?Units
Radiant flux Total radiation power in the specified beam or collection
Radiant intensity Power per outgoing solid angle
Radiance Power per projected area and per solid angle
Irradiance Incident power per receiving area
Radiant exitance Outward emitted power per emitting area
Resolving wavelength or frequency adds a further “per unit interval”, as in spectral radiance. The notation is consistent with NIST's table of spectral quantities. In particular, radiant flux is power, whereas the existing astronomical radiative flux is power per area. [ref].
This bookkeeping makes an optical camera's collected signal calculable from a source's radiance and the geometry of the receiving aperture. It also lets radiation measurements from different instruments be compared. For example, NIST describes separate calibrations of spectral radiance and spectral irradiance for instruments used to observe Earth. [ref].
The study of light propagation, absorption and scattering in marine water. It connects spectral measurements to dissolved material, suspended particles and biological activity.
The illuminated region of a water body. In ecological usage its photosynthetically usable part is often identified with the euphotic zone; the operational light threshold must be stated.
The upper water interval with sufficient light for substantial photosynthesis. A common operational lower boundary is the euphotic depth at 1% of surface photosynthetically active radiation.
The depth of a chosen light threshold, commonly 1% of just-below-surface photosynthetically active radiation. It depends on spectral attenuation and is not a universal species-specific compensation depth.
The spectral distribution of light emerging from a water body, determined by absorption and scattering by water, dissolved material, particles and sometimes the bottom. It supports remote inference of biological and optical properties.
The logarithmic rate at which downwelling spectral irradiance decreases with depth. It depends on absorption, scattering and the angular light field, rather than equaling the absorption coefficient.
Inference of chlorophyll a concentration from measured optical signals. Simple blue/green ratios rely on assumptions about covarying absorbers and scatterers; coastal mixtures can violate those assumptions.
The rate per unit path length at which light is scattered into the backward hemisphere. It influences water-leaving radiance and must be distinguished from the absorption coefficient.
The ratio of water-leaving radiance to incident downwelling irradiance just above the surface. It depends on both absorption and backscattering and has units of inverse steradian.
An optically defined non-phytoplankton particle category, including detritus and suspended mineral material. Its absorption is often strongest in the blue; scattering can also strongly affect remote-sensing reflectance.
The light-absorbing fraction of dissolved organic material. Its visible absorption commonly decreases approximately exponentially from blue to red, and can confound chlorophyll retrieval.
Solar radiant power incident per unit area. Horizontal, beam-normal and tilted-plane irradiances are different quantities: for a direct beam on a horizontal plane, , where is solar elevation.
Inferring properties of a distant object from measured signals, often electromagnetic radiation. Spatial sampling, spectral information and observation cadence must jointly resolve the physical process of interest. Received radiance alone need not uniquely separate temperature, emissivity and atmospheric attenuation.
Range measurement using transmitted laser light and its returned signal. For a pulse traveling a round trip in air, the elementary range is , where is the propagation speed and the measured travel time. Sensor position, pointing and time synchronization place returns in a three-dimensional point cloud. Scanning lidar can record multiple returns from vegetation and the ground; a return is an observation of a reflecting surface, not automatically bare-earth elevation.
Lidar acquired from an aircraft with surveyed trajectory and orientation. Ground-return classification removes canopy and structure returns before constructing a bare-ground digital elevation model. Penetration through forest is sampling through canopy gaps, not unrestricted transmission through leaves. Areas with few or no ground returns require uncertainty assessment rather than treating interpolated heights as observations.[ref]
Recovery of object positions and shape from photographs with overlapping views. Known or estimated camera geometry and matched image features constrain three-dimensional point positions. A metric reconstruction requires scale and a coordinate reference; image pixel size is not automatically the positional accuracy of the reconstructed surface. Stereo photogrammetry and structure-from-motion photogrammetry use related geometry with different acquisition and camera-estimation strategies.
An identifiable image feature or target with independently surveyed coordinates, used to georeference and constrain a photogrammetry model. A check point is withheld from the fit to assess reconstruction accuracy. Distribution across the site, accurate survey and consistent height datums matter as well as the number of points.
A photogrammetry workflow jointly estimating camera geometry and scene structure from many overlapping images, commonly followed by dense surface reconstruction. Ground control points or accurately surveyed camera positions establish scale and georeferencing. Independent check points test accuracy rather than participating in the fit. A fine image ground-sampling interval does not alone establish equal vertical accuracy.
Three-dimensional reconstruction using two or more overlapping photographs from different viewing positions. Corresponding image points define viewing directions whose intersection estimates the surface position. Surveyed camera positions and orientations, camera calibration and ground control points establish the coordinate geometry. Ordinary optical stereo models the visible surface, which can be forest canopy rather than bare ground.
Imaging in many narrow, usually adjacent wavelength bands. Relative to broad-band multispectral imaging, it can resolve narrower mineral absorption features in spectral reflectance, helping distinguish surface lithologies, at the cost of data volume and often reduced signal per band.
Removal or modeling of atmospheric absorption and scattering contributions to a measured signal to estimate surface or water-leaving optical quantities.
The Geostationary Operational Environmental Satellite series, used for frequent weather observations from geostationary orbits.
A European meteorological satellite series including geostationary orbit missions for repeated regional weather observations.
A European optical multispectral imaging mission using Sun-synchronous orbits for repeated observations of land and coastal areas.
A series of Earth-observation missions using repeated optical and multispectral imaging from near-polar Sun-synchronous orbits.
Imaging in several wavelength bands. Differences in spectral reflectance can distinguish surface materials, including vegetation and water.
An image formed by dividing corresponding calibrated pixels in two spectral bands. If their signals have a common multiplicative factor, and , their ratio is and does not depend on . This explains suppression of some common illumination effects and enhancement of spectral contrasts. Additive atmospheric radiance and small denominators prevent exact cancellation and can amplify errors. Normalized difference vegetation index is a related normalized spectral contrast.
The normalized contrast of red and near-infrared reflectances. Large positive values indicate a vegetation-like spectrum, but soil, atmosphere, shadows and canopy saturation affect its interpretation.
A displayed image assigning measured bands to colour channels that differ from ordinary visual perception. Assigning near-infrared to red is a common way to highlight vegetation.
The steep increase of healthy green vegetation reflectance between red absorption and the high near-infrared scattering plateau, typically around 0.68–0.75 micrometres.
Comparison of coherent radar phases from repeated observations to infer relative line-of-sight displacement. A complete deformation fringe represents half a radar wavelength after other phase contributions have been removed.
An image of the relative radar phase between co-registered coherent observations. A displacement changes the two-way path length, giving phase change proportional to , with sign set by the adopted range convention. A full wrapped phase cycle therefore corresponds to half a radar wavelength of line-of-sight displacement. Orbital geometry, topography, atmospheric delay and scattering changes must be separated from deformation; a fringe is not uniquely an elevation contour or an earthquake displacement.[ref]
A normalized measure of consistency of complex radar returns over a neighborhood. Low coherence makes interferometric phase unreliable and can arise from vegetation, surface changes, viewing differences or extreme displacement gradients.
The component of a displacement vector along a specified sensor–target viewing direction. Its sign must state whether motion toward or away from the sensor is positive.
The phase difference of complex radar returns. It includes deformation, topography, orbital geometry and propagation contributions; the displayed wrapped phase is defined modulo .
The fraction of transmitted light scattered sufficiently far from the forward direction under a specified measurement geometry. A transparent electrode can have high total transmittance yet blur an image if scattering is strong. State the collection angles and wavelength when comparing haze measurements.
The photon flux incident from all propagation directions, resolved by wavelength when needed for photodissociation. It is distinct from the net radiative flux through a horizontal surface; scattered and diffuse photons can contribute even when their net vertical flux cancels.
For an optical beam, time-integrated light intensity, giving incident energy per unit area.
Energy carried by incident photons per unit area. For monoenergetic photons, , where is photon fluence. With charged-particle equilibrium, relates it to absorbed dose.
The number of photons incident per unit area in a specified geometry. It is distinct from a rate per unit time. For a monoenergetic beam, photon energy fluence is .
A radiometer measures electromagnetic radiation through its received power. A radio radiometer can express that power as a noise temperature, allowing an unknown source to be compared with calibrated reference temperatures.
For a stable linear radiometer, known cold and hot radiances give . Put and . Then ; all three measured voltages contribute to the retrieved uncertainty.
A noise temperature expresses received or internally added noise power on an equivalent temperature scale. For one matched receiving channel in the classical thermal limit, an ideal rectangular bandwidth gives , with Boltzmann constant . An equivalent noise temperature need not equal a component's physical temperature.
The system noise temperature includes all noise contributions referred to a specified receiver input. In a simple additive model with negligible intervening losses, , where is the antenna temperature and the receiver noise temperature.
The receiver noise temperature represents noise added by the receiver as an equivalent noise temperature referred to its input. Separating it from the antenna temperature prevents internal receiver noise from being mistaken for radiation from the sky.
For a stable radiometer with affine output , let be the fixed response coefficient and the receiver noise temperature. Two reference loads with known equivalent input noise temperatures give outputs and . Subtracting their outputs eliminates ; dividing source and reference differences eliminates :
This two-load radiometer calibration assumes unchanged response and receiver noise temperature between readings, with the loads and source referred to the same input. It also eliminates an additional constant output offset. Physical load temperatures must first be converted to the applicable equivalent input noise temperatures when the classical thermal approximation fails. The Penzias-Wilson microwave background experiment gives an original worked example.
The fraction of incident radiant flux returned by reflection, for specified wavelength, angle of incidence and collection directions.
A function relating reflected radiance in one direction to incident irradiance from another direction at a surface location. A local nonemissive, nonfluorescent optical reflection model integrates over incident solid angle to obtain outgoing radiance. Transmission, wavelength conversion and nonlocal transport need additional models.
The ratio of reflected to incident radiant flux as a function of wavelength, under specified geometrical and illumination conditions. Unlike a perceived colour, spectral reflectance describes a physical response; perceived colour also depends on the incident spectrum and the observer.
The transfer of energy from electromagnetic radiation to matter. Reflection redirects the incident electromagnetic radiation, and transmission lets it pass through; absorption transfers some of its energy to the material instead.
Radiant flux is the power carried by electromagnetic radiation: radiation energy transferred per unit time, measured in watts, . Specify what is counted: all radiation emitted by a source, radiation within one beam, or radiation intercepted by a photodetector. Writing for the cumulative energy crossing the specified boundary gives
The first expression is an instantaneous derivative; the second is an average over the stated time interval. If arrives during , the mean radiant flux is . The photodetector's absorbed power equals this incident radiant flux only if it absorbs all of the incident radiation.
Keeping radiant flux fixed while widening a beam reduces its irradiance, the incident power per receiving area. Radiance additionally resolves the propagation direction and has units ; an optical camera can collect different radiant fluxes from sources with the same radiance but different apparent solid angles. Luminous flux, measured in lumens, weights the optical spectrum by the human eye's standardized response instead of counting energy equally at every wavelength. Equal radiant fluxes need not give equal luminous fluxes.[ref]
Radiant power also differs from the optical power of an optical lens, which is reciprocal focal length in dioptres. Likewise, astronomical radiative flux is power per area. Stating the units prevents these shared words from hiding different physical quantities.
In geometrical optics, suppose radiation has a nonnegative radiance whose directional contributions add as power. Here is the actual propagation unit vector, is a solid angle element, and the collecting oriented surface has chosen unit normal . Assume the following integrals exist. The outward and inward radiant fluxes are
Both quantities are nonnegative; their difference is the signed net radiant flux. The dot product supplies the cosine of the angle with the unit normal, hence the projected area. For incoming radiation the outward unit normal gives a negative dot product; the minus sign makes the collected power positive. Measuring the incoming angle from a source-facing unit normal instead gives the familiar positive . The units check is .
Equivalently, the incoming surface integral is with local irradiance ; the emitted part of is with radiant exitance . For a small flat patch and a narrow accepted cone, nearly constant radiance and angle give . This approximation requires uniformity across the patch and cone. Direction-independent outgoing radiance over a whole hemisphere gives the different exact result , by Lambert's cosine law and the spherical coordinates calculation in radiant exitance. Replacing the cosine-weighted integral by the hemisphere's solid angle would double the answer.[ref]
These radiance formulas presume a description in additive directional power; they do not reconstruct arbitrary coherent wave interference from separate ray contributions. For general electromagnetic fields, the signed instantaneous power crossing an oriented surface is , using the Poynting vector ; optical measurements commonly average over many wave periods. A zero net radiant flux can conceal equal nonzero incoming and outgoing radiant fluxes.
Consider an ideal parallel circular beam with uniform irradiance inside radius , zero outside, and total radiant flux . A centred circular collector of radius , perpendicular to the light rays, intercepts
For and , and the collected radiant flux is . Expanding the beam to while keeping fixed divides irradiance and the same collector's radiant flux by four: and . A collector covering either entire beam still receives .
Tilting a flat collector of actual area by an angle to the beam direction gives if its entire projected area remains inside the uniform beam. At the small collector's radiant flux halves. A collector large enough to intercept the whole beam after tilting still receives : the footprint on its surface grows as and cancels the irradiance factor. This model ignores diffraction, beam-edge structure, and changes in absorption with incidence angle; it counts incident rather than absorbed power.
Figure 1.
Fixed radiant flux, different beam widths and collecting projected areas
. Original ideal geometry, not measurements. The beam cross-sections use the same centimetre scale and total radiant flux of ; both collectors have radius . The side view shows a collector's cross-section at : its projection perpendicular to the light rays has half the width, with depth unchanged. The stated half-power result requires the projected collector to stay inside the uniform beam. Details.
Spectral radiant flux resolves radiant flux by wavelength or frequency:
Its units are per metre of wavelength or per hertz of frequency. A density quoted per nanometre must be integrated with wavelength intervals in nanometres; its numerical value is times the per-metre density. The frequency and wavelength spectral radiance conversion applies equally to spectral radiant flux, through the change of variables formula:
with the speed of light in vacuum and expressed per metre. The Jacobian determinant preserves integrated power rather than the height or location of an optical spectrum's peak.
For two explicitly synthetic optical spectra, let and quote the following densities in :
The triangle under and the rectangle under both have area : . Their band radiant fluxes differ: – contains of , while – contains of . These are specified functions, not measured source spectra.
Figure 1.
Equal total radiant flux can hide different band radiant fluxes and detector signals
. Original synthetic optical spectra with exact total radiant flux each. Darker shading marks the stated 100 nm bands. The lower panel uses a stipulated linear spectral responsivity ; integrating it against the two spectral radiant fluxes gives dark-subtracted electric currents and . The response is illustrative and is not a calibration curve for a real photodiode. Details.
The dark-subtracted electric current of a linear photodiode is
using its spectral responsivity. The figure's stipulated gives and despite equal total radiant flux. Symmetry places 's mean wavelength at and 's at ; a linear averages to its value at that mean. A single electric current therefore cannot determine an unknown broadband radiant flux without additional spectral information or a sufficiently flat calibrated response.
An optical power meter makes the definition operational: collect a specified beam on a calibrated photodiode, measure its electric current, and infer incident radiant flux using spectral responsivity. In the manufacturer-authored Model 1830-R Series Optical Power Meters User's Manual by Newport Corporation, §5.5.1 specifies a concrete sequence: connect the detector and its calibration module, select watt mode and the source wavelength, block the beam and press ZERO, then unblock it and read the display. The stored background includes dark current and ambient light present during zeroing. For narrowband radiation,
The electric current difference divided by spectral responsivity has units . For a stipulated and difference , the inferred radiant flux is ; these numbers illustrate the inference and are not reported measurements. The entire beam must hit the active area, and §5.6 calls for correct wavelength, background subtraction, and operation below detector saturation. If an optical attenuator is fitted, §5.4 requires the calibrated detector–attenuator pair and the matching meter setting.
The calibration itself must connect electric current to independently known incident power. NIST's Spectroradiometric Detector Measurements: Ultraviolet, Visible, and Near-Infrared Detectors for Spectral Power by Thomas C. Larason and others describes this chain. A cryogenic radiometer, an electrical substitution radiometer, compares temperature changes from absorbed light with those from known Joule heating (§5.2). Transfer and working detectors carry that watt scale to a spectral comparator (§5.1). A source, monochromator, focusing optics, shutter, beam splitter, and monitor photodiode illuminate a standard and test detector in turn; calibrated current-to-voltage electronics and voltmeters provide the readout (§§3.2.2–3.2.4, 4.3.1). At each wavelength, subtract blocked-beam signals and normalize each detector signal by the simultaneously sampled monitor signal. Comparison with the standard yields the test detector's spectral responsivity. NIST uses two working standards and alignment normal to the beam (§6.1.1).
Controls distinguish a smaller collected radiant flux from a faulty conversion: block and unblock the beam to check the baseline; check that its footprint fits the active area; verify the calibrated optical attenuator configuration; and test proportional response within the specified operating range. Source drift faster than monitor sampling is not removed. NIST discusses finite spectral bandwidth, stray light, temperature, and spatial response variations (§§3.1.1, 6.2, 7.2). A broadband optical spectrum requires the weighted integral in spectral radiant flux, rather than dividing by an arbitrary single spectral responsivity. These limitations explain why the meter's displayed watts need a stated collection boundary, spectral assumptions, and calibration conditions.
Radiant intensity describes how a source distributes its radiant flux over directions. A small cone of solid angle carries approximately when varies little across it; integrating over all emitting directions recovers the total radiant flux. For a source with isotropy emitting over a full sphere, and a cone carries . A planar source satisfying Lambert's cosine law instead has direction-dependent radiant intensity, even though its radiance is direction-independent.
An ideal point emitter emits radiant flux from one location. It approximates a physical source when its emitting dimensions are small relative to the distances relevant to the calculation. With isotropy, its radiant intensity is for total radiant flux . An extended source generally requires integrating contributions from different emitting positions.
Radiance resolves both position and direction: it is radiant flux per projected area and per solid angle. For a surface element of area , and a small outgoing cone at angle to its outward unit normal,
Here locates the patch and is a unit vector along the outgoing direction. The factor is the patch's apparent area viewed along that direction. The same local definition applies to transmitted or received radiation, choosing the positive projected area for the crossing direction. [ref].
For a planar patch of area with spatially uniform, direction-independent radiance , integrating over the patch gives
Take and . The radiant intensities are and . A sufficiently narrow cone therefore carries approximately or , respectively. Both views have the same radiance : the oblique view presents half the projected area. This is Lambert's cosine law, not a loss of radiance with viewing angle.
Figure 1.
One uniform patch: halving projected area halves radiant intensity while preserving radiance
. The geometric panels show a cross-section of a rectangular patch with the same depth perpendicular to the page. Parallel projection onto a plane perpendicular to the viewing direction changes its width by . The graph gives for the stated patch and marks the exact and cases. Details.
Radiance is useful because it separates a source's local directional output from the geometry of an observation. Integrating over outgoing directions gives radiant exitance; integrating incoming radiance at a receiver gives irradiance. In the stationary vacuum setting of vacuum conservation of specific intensity, a resolved uniform source retains its radiance along each light ray, while the radiant flux received from the whole source can decrease as its apparent solid angle shrinks.
The projected area of a planar patch viewed from a direction is the area of its perpendicular shadow onto a plane normal to that direction. If and are its unit normal and viewing unit vector, then by the dot product. On the outward side, this is . For a rectangle tilted around one edge, the width perpendicular to that edge contracts by and the edge length stays fixed, proving the formula. At grazing view ; interpret the radiance definition by a limit rather than dividing by zero.
Spectral radiance is radiance per unit frequency or wavelength interval. Its astronomical synonym specific intensity is usually written or ; this is a spectral radiance, distinct from the radiant intensity in radiometry. For positive small intervals,
The two expressions refer to the same spectral slice, with the positive widths related through frequency and wavelength . For frequency measured in hertz, , the units are . For wavelength measured in metres, . Integrating over a spectral band gives its radiance, and then multiplying by the collecting projected area and accepted solid angle gives a small uniform beam's radiant flux. A spectral measurement therefore retains information that the band-integrated power alone loses.
For an ideal blackbody, Planck's law gives and at temperature , independent of outgoing direction. Its existing frequency and wavelength peaks of Planck's law illustrate why the coordinate matters. The simpler flat-band calculation below isolates that same change of variables without the thermal physics.
In vacuum, , where is the speed of light. Increasing wavelength decreases frequency, so the ordinary signed differentials satisfy . A spectral slice's positive contribution to radiance must be unchanged. The change of variables formula therefore gives the absolute Jacobian determinant
These formulas use per metre and per hertz. The factor has units , which converts the former density into the latter. If denotes the numerical density per nanometre in , its per-metre value is . Conversely, the per-nanometre value is times the per-metre value. Likewise, a density per terahertz is times its per-hertz value because .
For a concrete stipulated spectrum, let between and , and zero outside. This is exactly throughout the band. Using the exact International System of Units value , the corresponding frequency bounds are
The transformed spectral radiance is on this interval. At it is approximately , and at it is in the same units. These are respectively and per terahertz. Thus a flat wavelength density becomes a decreasing frequency density; relabelling the horizontal axis alone would miss both the shape and the units. [ref].
The equality of band radiances has a short exact proof:
The reversed bounds and the Jacobian determinant preserve the same spectral measure. Each half-band, – and –, carries , although the corresponding frequency widths are about and .
Figure 1.
A flat wavelength spectrum becomes curved in frequency, with equal band radiances
. Matching colours identify the same two spectral half-bands. The shaded areas are each in either coordinate. The right-hand density is plotted per terahertz; the longer-wavelength half-band occupies the lower-frequency, narrower interval. Both total shaded areas are exactly . Details.
A spectrum's peak is consequently a statement about a particular density and coordinate. The frequency and wavelength peaks of Planck's law differ for precisely this reason. Integrated radiance over a specified physical band supplies a comparison that survives the change of variables.
In a transparent, stationary vacuum with Euclidean geometry and observers at rest in the same frame, the radiative transfer equation has zero absorption and emission, so along a light ray. The same specific intensity reaches the observer without geometric dilution. More generally, is a Lorentz invariant; collisionless propagation preserves this ratio along a ray even with gravitational or cosmological redshift, rather than preserving at shifted frequencies. [ref].
The elementary geometry is instructive. Two facing surface elements separated by distance exchange spectral radiant flux
The receiver subtends at the source. Describing that same traveling beam at the receiver instead uses . Both descriptions have identical geometric factors, so conservation of its radiant flux gives equal specific intensities at the two ends. This compares the same propagation direction at both ends, not emission in the opposite direction.
A uniform source remains equally bright per resolved direction under these assumptions, but occupies a smaller apparent solid angle farther away. For a small source of projected area at distance and a small receiver facing it, irradiance is approximately . The radiant flux intercepted by receiver area is approximately . With , and , distances and give and , and received radiant fluxes of and . The inverse-square law concerns collected radiation from the whole source; remains at either distance.
Limb darkening is the decrease of emergent specific intensity toward the apparent edge of a stellar disc. A transit depth then depends on the intensity sampled by the planet as well as its projected area; the uniform-disc area ratio is only an approximation.
For a spherical star of radius , let be distance from the centre of its projected disk and let be the angle between the outward unit normal and the line of sight. The angular coordinate is
Disk centre has ; the apparent edge approaches . A resolved image measures specific intensity in each direction. Limb darkening means that this specific intensity itself falls toward the edge; the projected area factor used when collecting radiant flux is a separate geometric effect.
A simple explanation assumes a horizontally uniform plane-parallel atmosphere, a sufficiently deep optically thick medium, negligible scattering of light, and thermal absorption and emission in local thermodynamic equilibrium. Let be inward vertical optical depth at fixed frequency . An oblique light ray has line of sight optical depth , so it encounters the same attenuation at a smaller vertical optical depth. The formal solution of the radiative transfer equation gives
Here is the source function; under these assumptions it is the Planck function . If temperature increases inward, an oblique view preferentially samples cooler layers and hence lower specific intensity. [ref]
The weight is an exponential distribution in vertical optical depth, with expected value and standard deviation . Thus “seeing down to optical depth one” refers to the line of sight, and represents a broad sampling of layers rather than a sharp emitting surface. For a source function that is linear throughout the atmosphere, the Eddington-Barbier relation is exact:
This follows by integrating the constant and linear terms separately. A constant source function instead gives constant specific intensity, as in the blackbody limit of an isothermal atmosphere. An outward-rising temperature or important scattering of light changes the argument. Close enough to the actual edge, a plane-parallel atmosphere also loses validity because of curvature. Neither one limb-darkening coefficient nor a universal dependence on wavelength follows from this simple model.
The linear limb-darkening law parametrizes angular specific intensity by a stipulated limb-darkening coefficient :
Here denotes specific intensity at one frequency, or radiance integrated over a specified interval of frequency or wavelength; it does not denote radiant intensity. Within this stipulated range, gives a uniform disk and . The linear limb-darkening law is linear in , while its radial profile is curved. In the linear source function example above, when and .
For the illustrative choice , the specific intensity at , , and the limiting edge is respectively , , and . These are synthetic values, not measurements of the Sun. Real optical spectra and stellar atmospheres need not obey a linear limb-darkening law; more flexible profiles can be necessary for accurate exoplanet transit photometry. [ref]
Figure 1.
Synthetic stellar disk and angular and radial limb darkening profiles
. The disk uses the stipulated linear limb-darkening law with . The profiles compare four stipulated limb-darkening coefficients at equal central specific intensity. Shading represents relative brightness, without prescribing a wavelength or copying a solar measurement. Details.
The parameter in the linear limb-darkening law specifies its contrast between disk centre and the limiting edge. Its interpretation requires the chosen law and interval of frequency or wavelength. A limb-darkening coefficient fitted to an exoplanet transit may trade off against the transit impact parameter and planet radius; it is not a direct measurement of one layer's temperature.
For a circular stellar disk whose specific intensity is independent of azimuth, its mean specific intensity over projected area is
The polar coordinates element is after angular integration, and . Substitution of the linear limb-darkening law gives
For a distant observer at distance receiving light normally to the detecting plane, the observed radiative flux is
It is spectral irradiance, with units when is per hertz. The local outward radiative flux per actual surface area is instead , a spectral radiant exitance. In a spherical model with the same stellar atmosphere everywhere, the luminosity is and , consistent with the inverse-square law. Compare disks at fixed or fixed total observed radiative flux explicitly: increasing lowers the total in the first comparison, while the second requires a larger .
A dark, opaque circular planet of radius blocks specific intensity over its overlap with the projected star. With and no transit dilution, the instantaneous exoplanet transit depth is exactly
When , the entire planetary disk is inside the stellar disk, and specific intensity varies little across that disk, small-planet transit weighting reduces to
Here is the projected centre separation divided by . The numerator samples the local specific intensity; the denominator normalizes by the whole star's mean specific intensity. The small-size argument and the distinction between full overlap and ingress are explicit in the original analytic treatment. [ref]
For a short approximately straight crossing at constant transverse velocity , write with at mid-transit, and let be the transit impact parameter. Then and . With stipulated and , a central crossing has mid-transit exoplanet transit depth approximately , whereas gives . A uniform disk would give for either fully overlapping crossing. The brighter central chord produces a deeper dip and the off-centre chord a shorter crossing at the same .
Figure 1.
Two transit chords across the same synthetic limb-darkened disk flux distribution
. Stipulated and ; the coloured curves integrate the occulted specific intensity over the finite planetary disk, including ingress and egress. Dashed curves show small-planet transit weighting only where the whole planetary disk is inside the star. Horizontal dotted segments give the uniform-disk full-overlap depth . The chord panel shows mid-transit positions; the horizontal coordinate in the light curve is distance travelled divided by stellar radius, equivalent to scaled time at constant transverse velocity. Details.
At fixed away from the edge, a two-dimensional Taylor's theorem expansion over the symmetric planetary disk cancels the first derivatives; the first correction to the blocked radiant flux is of order if the second derivatives remain bounded. This error estimate is not uniform at the edge, where the radial derivative of the linear limb-darkening law diverges. Ingress, egress, a grazing exoplanet transit, and a crossing of a starspot require the overlap integral or a suitable more complete model. Inferring without the local-to-mean specific intensity ratio can bias the estimated planet radius.
A motivating measurement used the McMath solar optical telescope at Kitt Peak. The Sun's image drifted across a -high, -wide optical slit; a double-pass spectrograph selected wavelength intervals from wide at to at . An EMI 9750 QB photodetector or silicon photodiode supplied a digitized signal: samples were summed into 4096 positions per crossing, including dark and off-limb sky readings. [ref]
Controls included daily realignment of optical telescope mirrors, wavelength settings between spectral lines insensitive to the rotational Doppler shift, selection of detectors without detectable hysteresis, and measured electronic nonlinearity corrections below of maximum signal. Limb-position calibration combined centres of gravity of the edge derivative profiles with almanac semidiameters; the preliminary edge shifts were consistent with astronomical seeing of approximately arcseconds full width at half maximum. [ref]
The decreasing centre-to-limb readout gives a specific intensity profile, constraining stratification of the photosphere under a radiative transfer model. Astronomical seeing, instrumental scattering of light, and solar granulation limit this inference; brightness alone does not uniquely determine temperature versus geometric height. These observations need not follow the stipulated linear limb-darkening law.
Radiant exitance is total outward emitted radiant flux per actual emitting area, measured in . Integrate radiance over the outward hemisphere, keeping the projected area factor:
If the radiance is direction-independent, the spherical coordinates element is , so
The weighted solid angle is , although the whole hemisphere covers . Directions near grazing contribute less because of the projected area. This geometric factor is the bridge between Lambert's cosine law and the total output of a diffuse patch. [ref].
For the uniform patch with , radiant exitance is . Its area gives total radiant flux over the outward hemisphere. This total differs from its radiant intensity , which describes only one direction.
For an ideal blackbody, integrate Planck's law over frequency as well: by the Stefan–Boltzmann law. An ideal emitting patch of area emits about outward. Irradiance counts incoming radiation separately; for purely emitted output the net outward exchange per area is . Reflected output must also be included if the surface reflects incident radiation.
Radiant power leaving a surface per surface area and per unit frequency or wavelength. Isotropic cavity radiation obeys ; the factor combines the outward hemisphere and projected area.
Irradiance measures incident radiant flux per actual receiving area. Its units match radiant exitance, but it counts the incoming side of the radiation exchange. Measure incoming angles from the unit normal pointing toward the incident hemisphere. Then
For distributed uniformly over , . If the receiving plane is tilted in the same broad uniform parallel beam, the incident power per actual area is halved to . This receiving-surface cosine factor is geometric and does not assume Lambertian reflectance. Together, irradiance and radiant exitance describe how radiation supplies and removes energy at a surface, while radiance preserves the directional information.
Incident radiant flux per unit area and unit wavelength or frequency. Integrating over a wavelength interval gives the contribution to irradiance in that interval. Conversion between and requires the absolute Jacobian determinant associated with , as for spectral radiance.
A parallel beam incident at elevation has horizontal irradiance . Its path optical depth in a plane-parallel atmosphere is . Confusing these projections changes both attenuation and intercepted power.
Radiometric absorptance is the fraction of incident radiant flux converted to heat in the specified receiver: , with for a passive receiver. Specify the entrance boundary, wavelength, incidence angle and receiver temperature. Radiation reflected or transmitted out of that receiver is not absorbed power. A cavity can improve radiometric absorptance by giving light repeated opportunities for wave absorption; a dark coating alone does not establish .
For broadband light, spectral radiant flux matters: . A single effective radiometric absorptance therefore depends on the incident optical spectrum as well as on the receiver. If the stated incident boundary lies before a window of known power transmission , the receiver absorbs when other optical losses are absent. Window loss and receiver radiometric absorptance are separate corrections.
An electrical substitution radiometer measures radiant flux by replacing absorbed optical power with calibrated Joule heating. At fixed temperature, feedback control holds the receiver equally warm with the beam blocked or admitted: after corrections, absorbed optical power equals the reduction in supplied electrical heating. Incident radiant flux additionally requires radiometric absorptance and entrance-path corrections. This connects radiometry to electrical standards.[ref]
Picture a small absorbing cavity connected to a colder reference block by a radiometer thermal link. A heater supplies Joule heating, a thermometer senses the receiver, and feedback control adjusts the heater. When light supplies some of the heating, the controller supplies less. The readout is the change in heater power, rather than the beam's brightness or its number of photons. NIST's institutional description gives a concrete receiver with a conical interior, absorbing coating and weak Kapton support, operated with a receiver heater and thermometer.[ref]
Figure 1. . Original functional schematic, not a drawing of a particular laboratory instrument. The shutter defines dark and illuminated states; incident radiant flux is specified at the receiver entrance. The cavity absorbs a fraction , while the remainder escapes. The reference block and receiver have distinct controlled temperatures. The thermometer signal drives feedback control, and heater voltage and electric current provide the electrical power readout. Lead heating, stray radiation and spatial heating differences require separate checks. Details.
An explicit lumped model makes the cancellation understandable. Treat the receiver as having uniform temperature , heat capacity , heater power deposited in the receiver, net parasitic input , and outward heat transfer to a reference at temperature . With incident radiant flux and radiometric absorptance , conservation of energy gives
The dark state has no admitted beam. Compare settled dark and light states, denoted and ; each has . Subtracting their balances gives
where and . At identical and , with unchanged surroundings and parasitic heating, both correction differences vanish. Thus , and . Notice the signs: extra stray thermal radiation admitted with the beam also reduces heater power, so it must be subtracted from that reduction before attributing the result to the intended beam.
Measure the heater's own terminal voltage and electric current : for steady electrical heating, , where is its electrical resistance. A supply voltage that includes wire voltage drops counts Joule heating in the leads too; only the portion deposited in the receiver belongs in this balance. Changes in that deposited lead heat belong in . Optical-electrical heating non-equivalence adds a further instrument-specific correction when the uniform-temperature model is insufficient.[ref]
A radiometer thermal link is the intentional path for thermal conduction from receiver to reference block. Near a chosen operating point, write with local conductance in . A weaker link increases the steady temperature rise per added power, , but also increases the uncontrolled relaxation time when the link dominates other heat transfer. These are local linear approximations: actual thermal conductivity depends on temperature, and active feedback control introduces its own response. NIST explicitly describes the tradeoff between weaker links and smaller receiver heat capacity.[ref]
Stipulate a model receiver held at with a reference at . Choose , so outward thermal conduction is . Ignore other heat transfer, take in both states, and stipulate perfect optical/electrical thermal equivalence. These values specify an original example; they are not measurements or claimed performance of a NIST instrument.
With the beam blocked, supply . With light admitted, the required heater power is . For stipulated radiometric absorptance at the chosen single wavelength,
The illuminated receiver still gets of heating. The remaining of incident radiant flux escapes without heating it. Calling the measured heater reduction “incident power” would underestimate the stipulated incident radiant flux by .
Figure 1.
The same receiver temperature with different electrical and optical heating
. Original stipulated steady-state example, not experimental data. Both columns total of deposited heating and balance the same outward thermal conduction. Incident radiant flux is , of which is absorbed; the that escapes is outside the heating column. No response time or measured uncertainty is implied. Details.
Two variations expose different pitfalls. First, if admitting the beam also adds of stray heating while the same is absorbed from the intended beam, the heater falls to . The apparent reduction is now ; subtracting recovers the beam's absorbed power. Second, keep the original heater readings but stipulate a light-state receiver offset at fixed reference temperature. The link carries an extra ; the inferred absorbed power becomes , and incident radiant flux becomes . This second variation deliberately violates exact fixed-temperature substitution.
The heater can supply only nonnegative power in this model. Consequently an absorbed beam above the available dark heater power cannot be compensated at the chosen setpoint merely by reducing the heater. The operating point, link, or admitted radiant flux must change; zero heater demand is a range boundary, not evidence of greater radiant flux sensitivity.
Optical-electrical heating non-equivalence means that equal absorbed optical and electrical powers do not necessarily give identical thermometer responses. Light and the heater may deposit heat in different places, producing different temperature gradients and different routes of thermal conduction to the support. Holding one thermometer reading fixed need not hold every part of the cavity at the same temperature. This differs from incomplete radiometric absorptance: one concerns where absorbed heat goes, the other how much incident light becomes absorbed heat.
A useful accounting convention is
where the signed correction is determined by characterization of the instrument and represents the residual departure from the lumped electrical substitution heat balance. It is not a universal extra source of energy. Varying the beam location or heater location, comparing equivalent temperature excursions, and checking repeatable dark returns can test the assumptions; no single small correction proves that every other correction vanishes.
A cryogenic radiometer uses low-temperature operation to improve electrical substitution radiometry through particular material and thermal properties. Lower heat capacity of a copper receiver permits a substantial cavity without proportionately slow relaxation; cold shields reduce unwanted thermal radiation, and heater leads with superconductivity can suppress resistive lead heating. These features do not make radiometric absorptance exactly one or abolish optical-electrical heating non-equivalence.[ref]
Cold surroundings matter as much as a cold receiver. In a simplified gray-surface model, net outward thermal radiation is , using emissivity , area , shield temperature and the Stefan-Boltzmann constant . Cold shields reduce the incoming term as well as the receiver's emission; cooling the receiver alone while leaving it exposed to warm surroundings does not remove incoming radiation. An entrance window still provides an optical and thermal path. Vacuum reduces gas-mediated heat transfer, while an intentional radiometer thermal link remains necessary. This model explains specific losses, rather than promising a universal improvement factor or perfect accuracy.
In NIST's 2001 comparison, the Laser-Optimized Cryogenic Radiometer held its cavity near and controlled the heat sink separately. Opening the beam reduced heater power at fixed cavity temperature; an AC bridge read the temperature. Tests used stabilized, spatially filtered lasers at , and , with horizontal input and a movable photodiode transfer detector in the cavity's optical plane. The paper contrasts this with the High Accuracy Cryogenic Radiometer's optical heating followed by two electrical heating cycles and interpolation.[ref]
The two laboratories compared calibrations of the same transfer detectors, and checked their spatial response. Alignment differences and transported dust limited that comparison; its agreement did not prove universal equivalence. Its correction tables retain window transmission, cavity radiometric absorptance, alignment, electrical readout and optical-electrical heating non-equivalence. The heat sink was controlled, but §3 does not specify its numerical setpoint: the reference in the preceding example is stipulated.[ref]
For a reproducible substitution measurement, record the beam's wavelength and entrance boundary, receiver and reference temperatures, heater terminal voltage and electric current, dark readings before and after illumination, and the settling rule. Check beam collection at the aperture, window transmission, radiometric absorptance, deposited lead heating, and sensitivity to beam position. Repeatability alone misses a stable correction error. Photometric calibration can subsequently use this radiant flux scale, but photometry still needs the appropriate spectral visual weighting; the cold receiver does not itself measure luminous flux.
Transmittance is the fraction of incident radiant flux transmitted through a specified object or boundary. Specify wavelength, angle of incidence, electromagnetic polarization and which outgoing light is collected.

Articles by others on the same topic (1)

Radiometry is the science of measuring optical radiation, which includes electromagnetic radiation in the visible and invisible spectrum, such as ultraviolet and infrared light. It encompasses various concepts and units related to the detection and quantification of light energy, including its intensity, power, and distribution across different wavelengths. Key terms in radiometry include: 1. **Radiant Power (or Radiant Flux)**: The total amount of energy emitted by a light source per unit time, measured in watts (W).