Transducer sensitivity describes how much the output of a measuring transducer changes when its input changes. For an output responding smoothly to input , a local model uses
VIM defines sensitivity of a measuring system through the quotient of corresponding indication and input changes, noting that sensitivity can depend on the input value and that the changes used must be appreciable relative to resolution.[ref] The derivative is the smooth local limit of that response ratio.
For an original pressure-to-potential transducer, stipulate
Its transducer sensitivity is
At , the output is and the sensitivity is . Across the finite interval from to , output changes by , so the measured response ratio is instead . Specify the operating point or interval rather than treating these two slopes as contradictory.
An output increment corresponds locally to input-referred resolution at the first operating point. This conversion is not a complete uncertainty statement. Increasing electronic gain can enlarge output noise as well as the signal; detector saturation can flatten the response and destroy local invertibility.
Estimate transducer sensitivity with known input steps, repeated outputs and checks for changing zero, loading and temperature. In a linear response , a constant offset cancels from differences; a changing offset does not. Retain both the sensitivity relation and its uncertainty in measurement calibration, and propagate the relevant contributions when reporting a measurement result.
Input-referred resolution expresses an output increment or distinguishable output change in units of the quantity applied to a measuring system. For local response with nonzero transducer sensitivity , an output increment corresponds approximately to
State what output criterion is being referred: one converter-code step, a noise-based threshold and a repeatably distinguishable input change are different specifications. Walt Kester's converter tutorial distinguishes code transitions, input-referred noise and noise-based resolution measures.[ref]
For an original pressure readout, suppose a transducer has , followed by an ideal 12-bit analog-to-digital converter spanning . Its code step is , corresponding to . This arithmetic specifies a nominal increment, rather than demonstrating that two pressures separated by that amount can be reliably distinguished.
Place an ideal amplifier of gain ten before the converter. The combined sensitivity becomes , so the converter step corresponds to . With zero offset in this model, the pressure span mapping onto the converter simultaneously shrinks from to . Gain trades converter coverage against nominal input increments; it does not create an unlimited input range.
Suppose the transducer itself contributes rms output noise. That is input-referred noise. The added ideal gain scales both that signal and that noise, leaving this pressure-noise contribution unchanged. More display digits therefore need not improve the useful measurement resolution.
For a nonlinear response, repeat the conversion at the relevant operating point and account for the actual observation bandwidth and processing. Near detector saturation, sensitivity can approach zero and the local conversion becomes uninformative. A complete measurement result also includes measurement uncertainty, which contains calibration and other contributions beyond the chosen resolution criterion.