Reuse the velocity selector analysis in the Thomson charge-to-mass experiment, but now send positive ions with into a separate magnetic analyzer. The selector fields and give . If the analyzer field is perpendicular to the selected velocity, its gyroradius gives
for a semicircular return to the entrance plane. The displayed impact separation is a diameter. This formula is linear in at fixed selected speed; the fixed-electric potential difference model above instead gives . They impose different experimental conditions.
For a chosen example, take , , and . Thus . Approximating an ion mass by and setting gives and . Here denotes the unified atomic mass unit. These are stipulated teaching inputs, not a reported Aston measurement or an exact magnesium atomic mass.
A finite accepted speed range broadens the radius distribution. At fixed and , . At fixed electric charge and mass, gives for a small kinetic energy spread. A one-percent kinetic energy spread therefore produces roughly a half-percent radius spread before other broadening is included. Its width convention must be specified before converting that statement into a mass resolving power.