Southern Illinois University's ET 304A guide, Lab 4: Voltage Measurements and Meter Loading Effects, documents an instructive test: build equal-arm voltage dividers on a supply, then measure across the lower arm with a digital multimeter, analog multimeter, and oscilloscope. Starting with two resistors, repeat with equal pairs from to . It requires separate recorded readings, unloaded predictions, and predictions including each instrument's input electrical resistance. The guide obtains meter resistances from specification sheets and scope resistance from the input-panel marking. Equal ratios control the unloaded target while changing susceptibility to voltmeter loading. The provided tables are blank: the guide specifies a procedure, not measured outcomes.[ref]
Preparation and accuracy checks are made concrete by Berkeley's Lab 1: Introductory Experiments and Linear Circuits I. Its different divider uses nominal and resistors and a supply on a breadboard. Measure isolated resistors first, record actual supply voltage, recalculate with these measured values, then record the junction voltage with the digital multimeter. It explicitly compares actual values with resistor tolerance and the final voltage with a range-dependent accuracy bound from the instrument manual. This is an uncertainty check, not an independent measurement calibration.[ref]
For a controlled comparison, additionally record the selected instrument range, probe factor, and input coupling; connect each instrument separately to the same electrical reference potential, and monitor the input rather than treating the supply label as an exact voltage. These are proposed controls for interpreting the documented procedures. A voltage indication and a resistance indication from the same digital multimeter do not independently validate its measurement calibration; a traceable reference comparison would be needed for that. Resistor tolerance describes component variation, while measurement uncertainty describes what is known about measured values. Using nominal arms alone can confuse mismatch with loading: equal nominal arms each within one-percent tolerance can give unloaded ratios from to .
The inference to test is specific: once actual arm values, input voltage, and meter input electrical resistance are accounted for, the readout should approach the loaded voltage divider prediction within the remaining measurement uncertainty. A discrepancy can instead reflect contacts, changing temperature, source droop, or an input that is not a fixed passive resistor. A source with positive series electrical resistance can be modeled by replacing with only when denotes its ideal internal voltage; when the actual voltage across the divider is measured, that terminal value is already . During settling, or with appreciable input capacitance and changing signals, the steady direct current calculation is insufficient. Neither institutional procedure supplies measured values for the original example above.

There are no discussions about this article yet.

Articles by others on the same topic (NaN)

There are currently no matching articles.