Voltmeter loading is the change in an electric potential difference caused by the electric current drawn by the measuring voltmeter.
For a loaded voltage divider in steady direct current, a voltmeter connected across the lower resistor places its finite input electrical resistance in parallel with the lower arm's electrical resistance and any intended electrical load of electrical resistance . With positive, fixed electrical resistances, an upper resistor of electrical resistance , and an ideal voltage source supplying electric potential difference , Kirchhoff's current law and Ohm's law giveHere is the connected circuit's electric potential difference across the voltmeter; omit if there is no intended electrical load. The parallel connection and need for large input electrical resistance are standard voltmeter operating principles.[ref]
10.4 Electrical Measuring Instruments. §10.4, Measuring Voltage with a Voltmeter: parallel connection and large input resistance
Voltmeter loading is a bias relative to the target electrical circuit with the voltmeter absent and any intended electrical load still present. Voltage divider loading error instead compares the attached-load output with the wholly unloaded output. An accurate readout of , supported by measurement calibration, can still differ from either target: calibration accuracy concerns the indication of the connected electric potential difference, whereas voltmeter loading changes that electric potential difference itself.