Photometric calibration 2026-10-14
NIST Measurement Services: Photometric Calibrations by Yuqin Zong and others describes a concrete experiment (§§3.2–3.3.2): a silicon photodiode, multilayer photometric matching filter, precision aperture, temperature sensor and current-to-voltage converter. A monochromator supplies a small beam; scanning wavelength yields spectral responsivity, and scanning position checks response across the aperture. Finite spectral bandwidth requires correction. The average aperture response converts uniform illumination to illuminance.
On the photometric bench (§§3.4.1, 3.5.3, 3.5.6), align lamp and aperture reference planes, set the lamp's electric current, stabilize, and compare a check lamp with standard photometers. The readout is a voltmeter signal. Subtract the shutter-closed reading (§3.4.4), divide by converter gain to infer electric current, and use calibrated illuminance responsivity to obtain . For a distant source, luminous intensity follows from , with between reference planes.
Controls include repeat lamp operations, distance changes to test the inverse-square law, temperature corrections and gain linearity checks. Spectral mismatch correction is source dependent (§3.4.3). Unknown optical spectra, spatial nonuniformity and incorrect reference planes remain limitations.[ref]
NIST Measurement Services: Photometric Calibrations by Yuqin Zong and others. §§3.2–3.5.6, printed pp. 16–19, 22–28, 33–37: apparatus, dark subtraction, readout, corrections and check-lamp procedure