A photometer measures a visually weighted radiation quantity. A filtered photodiode approximates the photopic luminous efficiency function with its spectral responsivity; calibration establishes how its electric current maps to illuminance or collected luminous flux. NIST's detector-based realization traces the spectral responsivity to a cryogenic radiometer and the aperture area to length measurements. [ref]
NIST Measurement Services: Photometric Calibrations by Yuqin Zong and others describes a concrete experiment (§§3.2–3.3.2): a silicon photodiode, multilayer photometric matching filter, precision aperture, temperature sensor and current-to-voltage converter. A monochromator supplies a small beam; scanning wavelength yields spectral responsivity, and scanning position checks response across the aperture. Finite spectral bandwidth requires correction. The average aperture response converts uniform illumination to illuminance.
On the photometric bench (§§3.4.1, 3.5.3, 3.5.6), align lamp and aperture reference planes, set the lamp's electric current, stabilize, and compare a check lamp with standard photometers. The readout is a voltmeter signal. Subtract the shutter-closed reading (§3.4.4), divide by converter gain to infer electric current, and use calibrated illuminance responsivity to obtain . For a distant source, luminous intensity follows from , with between reference planes.
Controls include repeat lamp operations, distance changes to test the inverse-square law, temperature corrections and gain linearity checks. Spectral mismatch correction is source dependent (§3.4.3). Unknown optical spectra, spatial nonuniformity and incorrect reference planes remain limitations.[ref]
For collected spectral radiant flux and a linear photodiode with spectral responsivity , the dark-subtracted electric current is . A reference optical spectrum sets the conversion from electric current to collected luminous flux. Applying it to another optical spectrum gives
Assume nonzero calibration and test signals. Then is the corrected luminous flux. Multiplying either optical spectrum by an overall positive scale leaves unchanged. If is proportional to , for every optical spectrum. Uniformly overfilled aperture measurements use wavelength-resolved irradiance instead, multiplying the electric current expression by aperture area; the same ratio cancellation gives the illuminance correction. This is the algebra behind source-dependent calibration, rather than a new measured correction curve. [ref]
Use the triangular in equal radiant power under visual weighting and stipulate the detector's spectral responsivity
This model has nonnegative response: is zero outside –. For a ideal calibration line at , and , so . An equal-radiant flux test line at gives , and . The unchanged conversion indicates , whereas the model's intended weighted result is . The required spectral mismatch correction is : the uncorrected result is 50% too large. All these numbers are declared model outputs, not instrument measurements or standard photometry data.
Figure 1.
Calibration at one wavelength leaves a spectral mismatch at another
. Original schematic and analytic model. The apparatus shows the signal path of a filtered photodiode photometer, with shutter-closed subtraction; it is not a drawing to scale of a specific instrument. The lower panels use the declared triangular weighting and detector response, not CIE tables. The two ideal lines carry the same collected radiant flux; the source-dependent factor corrects the test indication within this model only. Details.

Articles by others on the same topic (2)

A photometer is an instrument used to measure the intensity of light. It quantifies the amount of light that passes through or is emitted from a source, and it can be employed in various scientific and practical applications. Photometers can measure light in different contexts, such as luminous intensity (in terms of lumens), irradiance (in watts per square meter), or illuminance (in lux), depending on the specific type of photometer.