Photometer 2026-10-14
A photometer measures a visually weighted radiation quantity. A filtered photodiode approximates the photopic luminous efficiency function with its spectral responsivity; calibration establishes how its electric current maps to illuminance or collected luminous flux. NIST's detector-based realization traces the spectral responsivity to a cryogenic radiometer and the aperture area to length measurements. [ref]
For collected spectral radiant flux and a linear photodiode with spectral responsivity , the dark-subtracted electric current is . A reference optical spectrum sets the conversion from electric current to collected luminous flux. Applying it to another optical spectrum gives
Assume nonzero calibration and test signals. Then is the corrected luminous flux. Multiplying either optical spectrum by an overall positive scale leaves unchanged. If is proportional to , for every optical spectrum. Uniformly overfilled aperture measurements use wavelength-resolved irradiance instead, multiplying the electric current expression by aperture area; the same ratio cancellation gives the illuminance correction. This is the algebra behind source-dependent calibration, rather than a new measured correction curve. [ref]