Here the complex viscoelastic modulus is a stress-to-strain response coefficient, rather than a complex number's absolute value. Use strain and stress , with
The storage modulus measures the elastic component; the loss modulus measures dissipation. The mean power per volume is , and the energy dissipated per cycle is , fixing the sign convention.
A Hookean solid has , while a Newtonian fluid has . The Maxwell model interpolates with , giving and . It is fluid-like at low frequency and elastic on times short compared with relaxation time .
A rheometer imposes small-amplitude oscillatory shear and resolves the torque amplitude and phase relative to the imposed motion. The torsional measurement of a complex shear modulus makes the conversion from torque and twist to explicit. A frequency sweep locates relaxation scales; keeping the strain small tests the linear-response assumption.
A viscoelastic material stores part of the deformation energy and dissipates part of it. In linear viscoelasticity, a small imposed shear strain produces a stress determined by its history, . A shear step probes stress relaxation; a stress step probes creep compliance. A purely elastic solid has persistent stress after a held strain, while a Newtonian fluid relaxes that stress and creeps continuously under constant stress.
For a Maxwell model, a spring of shear modulus and a dashpot of viscosity are in series:
For the convention ,
The storage modulus dominates at short times or high frequencies; the loss modulus dominates at long times or low frequencies. The energy dissipated per cycle per volume is . A Kelvin-Voigt model places spring and dashpot in parallel and instead gives bounded creep : a simple illustration of a viscoelastic solid. Real materials can require a spectrum of relaxation times rather than one element.
A small-amplitude oscillatory rheometer measures stress amplitude and phase relative to imposed strain, giving and . First sweep strain amplitude to identify the linear range; control temperature, sample ageing, wall slip and instrument inertia. Frequency-dependent phase lag alone is insufficient if these apparatus effects are unresolved.
Figure 1.
Original normalized Maxwell relaxation and frequency response
. Details.
Use a rheometer, for example a cone-and-plate rheometer or a parallel-plate rheometer, to impose a small deformation and measure the required torque, or impose a torque and measure the deformation. The apparatus geometry converts these into shear stress and strain. First check that reducing the imposed amplitude does not change the normalized response; this identifies the linear regime and avoids yielding or structural rearrangement.
Three complementary protocols characterize the material's memory. In a stress relaxation experiment impose a small step and hold the deformation fixed. The output is the relaxation modulus , revealing how stored stress decays and whether an elastic plateau remains. In a creep compliance measurement impose a small stress step and record . Removing the load also distinguishes recoverable deformation from permanent flow.
In small-amplitude oscillatory shear, impose sinusoidal strain over a range of frequencies and measure both stress amplitude and phase. With , fit
The outputs are the storage modulus and loss modulus , or equivalently amplitude ratio and loss tangent . These measure elastic storage and dissipative response, respectively. The step experiments give time-domain information and the oscillatory experiment gives the corresponding frequency-domain response. All comparisons must use the same deformation convention and controlled temperature.
Rheology Created 2026-09-24 Updated 2026-10-14
Rheology studies how matter deforms and flows in response to applied stress.
Rheometer Created 2026-09-24 Updated 2026-10-14
A rheometer imposes a controlled deformation or stress and measures the material response to determine a constitutive equation.
Rheometry Created 2026-10-20 Updated 2026-10-25
Measurement of stress–deformation response using a calibrated rheometer. Geometry, temperature, sample history and possible wall slip must be controlled before interpreting a measured force or torque as a bulk constitutive response.