Thermocouple voltage calibration relates an actual assembled sensor's electromotive force to independently established junction temperatures. It must be distinguished from selecting a nominal letter-type conversion curve. In Thermocouples Calibrations Services by National Institute of Standards and Technology (opening method description and bath-calibration table), NIST describes comparison with a maintained reference thermocouple, comparison with a standard platinum resistance thermometer in controlled heat baths, and fixed-point calibration at the freezing points of zinc, aluminum, silver and gold on the International Temperature Scale of 1990. These are independent thermal references, unlike a test that checks only a voltage-input circuit.[ref]
For a concrete electrical check, the SCC-TC guide describes shorting the input after 30 minutes of powered operation and adjusting its offset to a zero-voltage readout. Its separate calibrator procedure requires equal connection temperatures at the calibrator and module, measures the supplied voltage, converts it to temperature, and compares with the calibrator setting. Such an injected-signal test checks the readout chain; it does not expose a real measuring junction to an independently known bath temperature.[ref]
Useful additional checks follow from the model: log the raw voltage, the reference temperature, and the final inferred temperature separately; hold stable while changing and test whether cold-junction compensation preserves the inferred ; compare against an independent thermometer at several stable heat bath settings. Check immersion and repeatability instead of assuming a steady display establishes thermal equilibrium. These are proposed diagnostic checks, not new experimental results. The equal-temperature and polarity controls are already described under Seebeck effect.
Limits arise separately from the thermocouple reference function, sensor-wire changes, thermal contact, terminal temperature differences, and the voltage-input circuit. In particular, the SCC-TC guide warns of connector-to-module temperature gradients and voltage drops caused by the fault-detection electric current flowing through lead electrical resistance.[ref] A fitted curve or compensation algorithm cannot make all these errors vanish. No single accuracy is assigned to every thermocouple.
A thermocouple reference function gives the nominal thermoelectric electromotive force versus temperature for a specified material pair and reference temperature, conventionally . Real Seebeck coefficients vary with temperature, so
is generally nonconstant. A lookup calibration can use interpolation between tabulated values; a fitted calibration can use polynomials over stated intervals. Use the correct material type, units, reference and validity interval in both the forward and inverse conversions. The FP-TC-120 manual specifically requires the channel's thermocouple type to be configured and describes its use of NIST-175 reference characteristics.[ref] A nominal reference curve is not evidence that an aged or inhomogeneous wire still follows it.
A local error estimate explains why the reference thermometer matters. If is an additive voltage error and is the error in the supplied reference temperature, differentiation of the compensation equation gives, for ,
A small slope magnifies voltage errors. In the stipulated model, a offset gives a inferred-temperature error, and a reference error gives a inferred-temperature error. These are sensitivities of the chosen model, not universal performance figures for real sensors.