Branch-current mismatch in a resistance ratio measurement is a residual scale error when the sensor and reference resistor carry unequal electric currents. A common supply or common nominal current setting does not establish the shared excitation required by a shared-excitation resistance ratio measurement.
Suppose separate branches carry positive currents and . Ideal potential sensing givesThus the fractional resistance error is exactly . If and , the common scale cancels, whileIn an unbranched series circuit the same electric current passes through both components; this particular error instead concerns separate paths, leakage, or unequal effective sampling times.
For an original example, use , , and . Their potentials are and . The inferred resistance is , high by approximately . Each current is only from its nominal setting, but the opposed errors add to first order.
A useful ideal comparison is current-source swapping in a resistance ratio measurement. In two otherwise unchanged acquisitions, assign source A to the sensor and source B to the reference, then exchange them. If , the two ratios obeyThis exact geometric mean cancellation assumes unchanged source outputs and resistances between acquisitions; switching transients and sensor self-heating in a ratiometric resistance measurement can violate that assumption. The primary circuit guide independently treats current-source mismatch and source interchange in its three-wire lead-cancellation circuit; that topology's correction must follow its own equations rather than be replaced by the two-branch model above.[ref]
A proposed test measures both branch currents through characterized sensing resistors, then changes their assignments while holding the sensor environment fixed. Agreement after swapping tests the mismatch model; agreement from merely changing the common nominal current tests a different cancellation.
Current-source swapping in a resistance ratio measurement exchanges two excitation sources between the sensor and reference resistor, allowing an unequal-current contribution to be identified or canceled. The combining rule depends on the circuit topology. In the separate-branch model of branch-current mismatch in a resistance ratio measurement, the exact combination is a geometric mean.
Let and let the two positive source currents have a fixed ratio . With source A on the sensor and B on the reference, the measured potential ratio is . After exchanging their assignments, . HenceThis assumes that the resistances, current outputs, potential-sensing gains and relevant thermal conditions are unchanged across the two acquisitions. It cancels the multiplicative source ratio, not an additive readout error or an inaccurate assigned reference resistance.
For an original example, let and . The acquisitions give and . Their geometric mean is exactly . Their arithmetic mean is , high by approximately . In general,The arithmetic mean suppresses a small mismatch to second order here, but is not the exact inverse-pair cancellation.
The primary RTD guide uses source interchange in a different three-wire lead-cancellation circuit. Its two readings are averaged according to that circuit's equations; they must not be replaced automatically by the geometric-mean rule above.[ref]
A drift residual after current-source swapping remains if the source ratios in the two acquisitions are and . For unchanged , the same calculation gives . A increase from to therefore makes the combined estimate low by . Sensor self-heating in a ratiometric resistance measurement can also change between source assignments.
A proposed apparatus check switches two characterized current sources between stable sensing resistors, waits for electrical settling, and compares both individual readings and their product. Reversing the order of the two acquisitions helps distinguish a repeatable source mismatch from drift during the sequence. This tests the stated circuit model without presuming that all source-swapping circuits use the same estimator.
Drift residual after current-source swapping 2026-11-05
A drift residual after current-source swapping remains when the current ratio changes between the two source assignments. The geometric mean cancellation in current-source swapping in a resistance ratio measurement is exact for a fixed multiplicative mismatch; it does not make a time-varying mismatch constant.
Let the true resistance ratio remain fixed. At the two acquisition times , write . With opposite source assignments,For , the logarithmic error is . A full-interval bound therefore givesFor an original example, a logarithmic source-ratio drift of over a acquisition separation produces . This is a underestimate for that drift direction. Shortening the separation reduces this particular residual, provided both readings still follow electrical settling.
A useful original logarithmic variant of an ABBA measurement sequence uses source-assignment signs at equally spaced times . The four readings obey . Combining them as givesBoth the constant and linear logarithmic drift terms cancel because the signed time sums vanish. Unequal timing, nonlinear drift and a changing sensor resistance can leave residuals. If itself varies, the combination includes its geometric average over the acquisitions rather than automatically recovering a single-time value.
The primary RTD guide establishes source interchange in its three-wire circuit; its topology-specific arithmetic averaging is distinct from the separate-branch logarithmic model above.[ref]
A Basic Guide to RTD Measurements by Joseph Wu. §1.4.2, printed p. 6, Figure 1-6 and equations (9)–(12)
A proposed bench test imposes a slow controlled source-ratio variation, timestamps the readings, and compares a two-reading swap with the balanced four-reading sequence while using stable sensing resistors. Joseph Wu (electrical engineer) 2026-11-05
Joseph Wu is the named author of Texas Instruments' 2018 guide to resistance thermometer measurements, revised in 2023. The guide connects sensor wiring, excitation electric current, reference resistors and analog-to-digital converters rather than treating the converter as a complete thermometer by itself.[ref]
A Basic Guide to RTD Measurements by Joseph Wu. Title page, revision record and §2, measurement circuits
One concrete design distinction is between two-wire resistance measurement and four-terminal sensing. In the two-wire arrangement, the same leads carry excitation electric current and contribute to the measured electric potential difference. Their electrical resistance is therefore counted with the sensor. Separate sense leads with high input impedance can exclude that series drop, although input leakage and other nonidealities still require attention.[ref]
A Basic Guide to RTD Measurements by Joseph Wu. §2.1.3, lead-wire error; §2.6, four-wire measurement
For an original ideal example, let a sensor have electrical resistance , with two leads carrying . Two-wire resistance measurement givesThe lead error remains even when a shared reference resistor cancels excitation variations in the ratiometric measurement. Ideal four-terminal sensing instead measures directly across the sensor. This illustrates the platinum resistance thermometer lead resistance error mechanism without assuming a particular resistance-to-temperature calibration.
A proposed bench comparison replaces the sensor with a stable calibrated resistor and adds known series resistors to its current leads. Comparing two-wire and four-terminal sensing readings then tests lead-drop rejection without changing sensor temperature. The calculation distinguishes the wiring error from sensor self-heating in a ratiometric resistance measurement, which changes the sensor's actual electrical resistance.
Ratiometric measurement 2026-11-05
A ratiometric measurement infers a quantity from a ratio of two measured signals chosen to share an unwanted multiplicative variation. The useful question is which inputs actually share that variation: a reference resistor error or a change in the sensor itself generally survives the cancellation.
In a shared-excitation resistance ratio measurement, a current source passes the same electric current through a sensor of electrical resistance and a reference resistor . Their electric potential differences are and . An ideal normalized analog-to-digital converter reading is thereforeThe current source need not have a perfectly constant output for this cancellation, provided both measurements sample the same electric current, the denominator is nonzero, and the converter stays within its operating ranges. This is a standard arrangement for a resistance thermometer.[ref]
A Basic Guide to RTD Measurements by Joseph Wu. §2.1.3, pp. 10–11, shared-current resistance measurement
For an original ideal example, take , and . Then , and . Increasing by gives and : the ratio remains . If the converter instead uses an independent fixed reference and the calculation assumes the original , the inferred electrical resistance becomes .
Shared excitation cancels in a resistance ratio; a fixed independent reference retains the excitation error. Original ideal calculation, not measured data.
Details. The reference-resistor error in ratiometric resistance measurement remains even with perfect shared excitation. If the actual reference resistor is but the calculation uses its nominal value, thenThus a reference error makes the nominal estimate of a true sensor about . Reference accuracy and its variation with temperature still matter.[ref]
Nor does the ratio remove sensor self-heating in a ratiometric resistance measurement. Joule heating dissipates , which is in the example. A temperature rise changes the actual electrical resistance of a resistance thermometer; the ratio then measures that changed resistance. Reducing the electric current or comparing several currents can reveal this physical effect.[ref]
The law of propagation of uncertainty makes the role of covariance explicit. For small perturbations of nonzero mean signals, the first-order measurement uncertainty of obeysThe contribution from a shared excitation fluctuation cancels through the covariance term. Treating the two signals as independent random variables would discard that cancellation. Other errors need their own inputs in the complete model.[ref]
For time-varying excitation, ratiometric cancellation with matched signal and reference filters requires matching dynamics as well as shared wiring. With constant resistances and identical linear time-invariant systems of impulse response , the filtered signals are and , so their ratio is still wherever defined. Here denotes convolution. Different filters or different sampling times can leave an excitation-dependent error.[ref]
RTD Ratiometric Measurements and Filtering Using the ADS1148 and ADS1248 Family of Devices by Luis Chioye. First-page abstract and low-pass filter design discussion, matching input and reference filters
A simple proposed bench test first replaces the sensor by a stable calibrated resistor, sweeps the electric current within the instrument limits, and records both electric potential differences and their ratio. That separates electrical cancellation from changing sensor temperature. Repeating with a resistance thermometer in a stable bath then tests Joule heating; changing the reference resistance or one filter tests the predicted failure modes. These are proposed checks of the model, rather than reported experimental results.
Reference resistor 2026-11-05
A reference resistor is a resistor whose assigned electrical resistance supplies the comparison value in a resistance-measurement circuit. Its value may come from measurement calibration rather than its nominal marking. The useful reference is its value under the stated temperature, loading and connection conditions, together with the associated measurement uncertainty.
In a shared-excitation resistance ratio measurement, the sensor and reference resistor carry the same electric current. Measuring their electric potential differences gives , so . A reference value error therefore sets a scale error even when excitation cancels perfectly. Reference-resistor error in ratiometric resistance measurement derives the exact relation and distinguishes a correction from its remaining measurement uncertainty.[ref]
The connections define which resistance is compared. In four-terminal sensing, two leads carry the electric current while a separate high-impedance pair senses the electric potential difference at the reference terminals. This avoids including the current-lead drops in the assigned reference value, within the sensing instrument's limitations.
For an original ideal example, a reference resistor carries through two current leads, each of electrical resistance . The reference element has a potential difference of , while a reading outside both leads is . If this latter reading is used as the denominator while the calculation assumes , a true sensor is inferred asSensing directly at the reference terminals removes this idealized lead contribution. It does not correct a wrong reference value. The circuit guide explicitly shows a Kelvin connection at the reference terminals for this purpose.[ref]
A small temperature change shifts the value approximately by , with temperature coefficient of resistance . Excitation also dissipates electric power ; a reference with a different temperature or load from its measurement calibration may require a correction or additional measurement uncertainty. A proposed check compares the reference before and after the measurement sequence under the same connection and loading conditions. Repeating the comparison at a lower electric current tests for a heating contribution. Stability of the observed ratio alone cannot certify the reference's assigned value.
Reference-resistor error in ratiometric resistance measurement is the error left when an instrument converts a measured resistance ratio using an inaccurate value for its reference resistor. Shared excitation removes one nuisance input; it does not establish the denominator's true electrical resistance.
Let be the actual reference resistor value and the value used in calculation. For an ideal simultaneous reading , the inferred sensor resistance isWriting gives the exact fractional error . The familiar is its first-order approximation. If a deterministic bound is , the inferred fractional error lies inA symmetric reference tolerance therefore produces a slightly asymmetric resistance-error interval.
For an original example, suppose the instrument uses while the actual reference resistor is . A true sensor is reported as , an error near . Multiplying the shared excitation by any common positive number leaves this error unchanged in the ideal model. Using a measured and corrected instead of its nominal marking can remove the known part, while the corrected value retains its own measurement uncertainty.
Temperature variation can produce the remaining error after measurement calibration. Locally, , where is the temperature coefficient of resistance. For example, a -parts-per-million-per-kelvin coefficient over contributes roughly parts per million; it survives excitation cancellation. The primary circuit guide explicitly requires reference accuracy and low drift, and separates these requirements from shared-current cancellation.[ref]
A Basic Guide to RTD Measurements by Joseph Wu. §2.1.3, printed pp. 10–11, reference accuracy and drift
For small random errors, applying the law of propagation of uncertainty to givesThe covariance term must follow how the reference value and reading were obtained; it cannot simply be assumed zero when inputs are shared.[ref]