End-expiratory hold 2026-11-05
An end-expiratory hold is an interruption of airflow at the end of exhalation during which airway pressure is observed while the respiratory system is occluded. In an appropriate passive, leak-free measurement it can assess total positive end-expiratory pressure, including intrinsic positive end-expiratory pressure. The set external pressure alone need not equal this measured total.[ref]
How to measure driving pressure by Hamilton Medical. End-expiratory hold and measurement conditions
The mechanical reason is easiest to see in one communicating compartment. With airway resistance , let relate airflow to the difference between alveolar pressure and measured airway pressure. Once flow vanishes and pressure has equilibrated through the communicating airway, this model gives . Zero flow removes the resistive pressure drop; it does not mean that either pressure is zero.
For stipulated pressures, an external setting of and an equilibrated total of differ by . If plateau airway pressure is , respiratory driving pressure is , whereas subtracting the external setting would give 18. These constructed values illustrate reference selection, not patient settings.
Occlusion changes the boundary condition rather than directly opening every lung region to the sensor. A compartment isolated by airway closure cannot be assumed to equilibrate with the measured opening. Active muscular effort, a leak or a pressure trace that has not settled also defeats the simple passive-compartment inference. The observed airway-pressure plateau and the inferred regional alveolar pressure must therefore remain distinct.
Exhalation 2026-11-05
Exhalation is the outward movement of gas from the lungs through the airways. During ordinary quiet breathing, relaxation after inhalation allows elastic recoil to reduce lung volume. Forceful expiration can additionally involve active muscles.[ref]
How the Lungs Work: What Breathing Does for the Body by the National Heart, Lung, and Blood Institute. Breathing out
The term identifies a flow phase, rather than one universal muscle-activation pattern.For an ideal passive compartment, let be the excess volume above equilibrium with the outlet pressure. With airway resistance and respiratory-system compliance ,An original example uses cm HO·s/L, L/cm HO and L. The time constant is 0.60 s, leaving about 0.221 L excess volume after one time constant. After three time constants, only about 0.030 L remains. These are model calculations, not recommended expiration times.
The flow-driving pressure difference is between alveolar pressure and the pressure at the open airway. Pleural pressure affects the distension and geometry of the lung and airway but is not interchangeable with either of those endpoint pressures. Active expiration can also narrow intrathoracic airways, so greater effort need not produce proportionally greater flow.[ref]
Pulmonary Function Testing by Warren M. Gold and Laura L. Koth. Figure 25-1 and Mechanical Properties of the Respiratory System
The linear decay model is therefore useful for isolating resistance, compliance and available time, while measured flow can reflect changing airway geometry, active effort or multiple compartments. Completing one expiratory interval is a mechanical event; removing a given amount of carbon dioxide additionally depends on the composition and origin of the expired gas.
Inhalation 2026-11-05
Inhalation is the inward movement of respiratory gas into the lungs. During ordinary spontaneous breathing, contraction of the respiratory diaphragm and contribution from external intercostal muscles expand the thoracic space; the resulting pressure relationships permit inward flow.[ref]
How the Lungs Work: What Breathing Does for the Body by the National Heart, Lung, and Blood Institute. Breathing in and breathing out
Positive-pressure ventilation can instead assist inward flow by raising pressure at the airway opening.The flow-driving difference is between airway-opening pressure and alveolar pressure. In a simple resistive model,where is airway resistance. For constructed gauge pressures and cm HO, with cm HO·s/L, inward flow is 0.5 L/s. These are illustrative pressures, not a measurement or a prescribed breathing pattern.
Transpulmonary pressure answers a separate question:with the pleural pressure. If the same example has cm HO, the distending difference is 5 cm HO. At a later no-flow instant, can equal atmospheric pressure while remains negative. The expanded lung can then have positive transpulmonary pressure without continued inward airflow.
It is consequently misleading to use any single negative intrathoracic pressure as a synonym for inspiration. Pleural pressure, alveolar pressure, the applied airway pressure and volumetric flow rate refer to different parts of the mechanical system. Inhalation names the flow phase; the relevant pressure differences explain how it is produced.
Lung compliance 2026-11-05
Lung compliance measures how much lung volume changes for a change in distending pressure. For a static pressure-volume relation, the local value isThe relevant pressure is transpulmonary pressure, not alveolar pressure or airway-opening pressure alone. The pleural pressure reference separates the lung's elastic behaviour from that of the surrounding chest wall.
OpenStax connects lung expansion to pressure changes, elastic recoil and pulmonary surfactant, which modifies surface-tension effects.[ref]
Anatomy and Physiology 2e by J. Gordon Betts and others. §22.3, The Process of Breathing: pressure relationships, surface tension and compliance
A compliant lung needs a smaller pressure increment for the same small volume increment, but that statement alone does not describe recoil, airway patency or overall respiratory function.A finite pressure-volume ratio is a secant slope, while is a local slope. In an original nonlinear example, set and when volume is measured in litres and pressure in centimetres of water. Local compliance is . It falls from at to at L. Across that entire one-litre interval, however, the secant compliance is in the same units. These are constructed values, not clinical reference ranges.
Static versus dynamic lung compliance distinguishes elastic behaviour from measurements during airflow. A simplified lumped dynamic model adds an airway resistance term proportional to volume-flow rate: pressure depends on both and . A pressure ratio obtained while air is moving can therefore differ from the static elastic slope. Inflation history and hysteresis also require the direction and volume interval to be specified.
In a 1959 primary study, researchers used an intra-esophageal balloon connected to a differential pressure transducer, with a mouthpiece pressure connection, to obtain pressure differences relevant to lung mechanics.[ref]
Airway Resistance and Lung Compliance by Ole Storstein and others. Methods, intra-esophageal balloon and Sanborn differential pressure transducer; Yale Journal of Biology and Medicine 31, 387–396
Such a surrogate pressure measurement has its own positioning and calibration assumptions. Spirometry supplies volume information; volume measurement alone does not establish lung compliance.Report the pressure references, volume range, flow conditions and measurement history alongside a compliance value. A single number is a property of the specified experiment and state, not an invariant description of every breath.
Quiet breathing chiefly uses active inspiration and passive expiration; deeper or more forceful breathing recruits additional muscles and changes their coordination. The vertical, anteroposterior and transverse dimensions of the thorax can all change, and different patterns need not represent different mechanisms of gas exchange.
The respiratory diaphragm is the principal inspiratory muscle. Its contraction lowers and flattens its domes, increases the thoracic vertical dimension and displaces abdominal contents so the relaxed abdominal wall moves outwards. Its lower-rib attachments also contribute to expansion of the costal margin. The phrenic nerves, chiefly C3–C5, provide its motor supply. The external intercostal muscles elevate and stabilize the ribs, while the interchondral or parasternal portions of the internal intercostal muscles also assist inspiration. It would be wrong to give every internal intercostal fibre an exclusively expiratory function.
Elevation of upper ribs and the sternum increases the anteroposterior dimension, the pump-handle component. Outward movement of the lateral portions of lower ribs increases transverse diameter, the bucket-handle component. These are components of oblique rib rotation rather than two mechanically isolated systems. Abdominal or diaphragmatic breathing makes descent of the respiratory diaphragm particularly conspicuous; costal breathing makes chest-wall expansion more conspicuous. Both contribute in ordinary respiration, with their proportions changing with posture, mechanical loading and individual anatomy.
As the thorax expands, pleural pressure becomes more negative and transpulmonary pressure increases. The lungs expand and alveolar pressure briefly falls below atmospheric pressure, admitting air. Relaxation of inspiratory muscles lets elastic recoil reverse the flow in quiet expiration. The abdominal wall and respiratory diaphragm return towards their resting arrangement without requiring active abdominal compression for each quiet breath.[ref]
22.3 The Process of Breathing. Mechanisms of Breathing; Pulmonary Ventilation: pressure changes and inspiratory versus expiratory activity
With deeper inspiration, the scalene muscles help stabilize or elevate the upper ribs, and the sternocleidomastoid muscle can elevate the sternum when the head is fixed. With the upper limbs or shoulder girdle fixed, muscles such as pectoralis minor and serratus anterior can assist expansion in greater effort. Recruitment is graded; an anatomical list of possible accessory muscles does not mean all are active in every ordinary breath.
Active expiration uses the abdominal muscles, including rectus, oblique and transversus components, to compress the abdominal contents and drive the respiratory diaphragm upwards. The interosseous portions of the internal intercostal muscles help depress the ribs. This is important in vigorous exercise and expulsive actions. At rest, elasticity supplies much of the expiratory work; during high demand, active muscle contraction adds to it.
The observable pattern also changes with the physiological task. Exercise increases breath depth and usually frequency. A sigh is an enlarged inspiration followed by expiration; yawning includes a prolonged deep inspiratory movement. Coughing coordinates an inspiration, closure of the glottis, pressure-generating expiratory effort and sudden glottic opening. Speech imposes a controlled prolonged expiration with laryngeal regulation. Swallowing briefly interrupts ordinary airflow. Thus respiratory observation should include phase, depth, frequency and coordination, not merely count which ribs move. Fixed claims that men must breathe abdominally and women must breathe costally overstate a variable tendency. Pregnancy or a change of posture can alter mechanical contributions without replacing the basic pressure mechanism.
Ordinary inspiration is active enlargement of the thorax; ordinary expiration is chiefly passive recoil after the inspiratory muscles relax. The movements increase and decrease the vertical, anteroposterior and transverse dimensions together. Their relative prominence varies with posture and the individual; there is no exclusive division into abdominal and costal kinds of breathing.
The principal inspiratory muscle is the respiratory diaphragm. Its peripheral fibres shorten towards the central tendon, lowering and flattening its domes. This increases the vertical dimension of the thorax and displaces the abdominal contents downwards and forwards, making the relaxed abdominal wall project. Its costal attachments also act on the lower ribs: with abdominal support opposing unlimited descent, shortening can lift and expand the lower costal margin. Thus the respiratory diaphragm does more than pull down a piston. Its motor supply is through the phrenic nerves, principally from cervical segments C3–C5.
The external intercostal muscles elevate the ribs and help keep the intercostal spaces from being drawn inwards as pleural pressure falls. The interchondral, or parasternal, portions of the internal intercostal muscles also contribute to inspiration. It would therefore be inaccurate to label every part of the internal intercostal muscles expiratory. Coordinated activity braces the rib cage so that the shortening of one layer moves the ribs rather than simply deforming the spaces between them.
Rotation about the posterior articulations raises the anterior ends of the upper ribs and the sternum, increasing the anteroposterior diameter: the usual pump-handle description. The more laterally directed portions of the lower ribs rise and move outwards, increasing the transverse diameter: the bucket-handle description. These are descriptions of components of an oblique movement, not two independent sets of hinges. The scalene muscles can stabilize and elevate the uppermost ribs, with recruitment varying in quiet breathing. The sternocleidomastoid muscle, when the head is fixed, is an accessory elevator of the sternum in greater inspiratory effort; it is not required for an ordinary quiet breath.
The lungs follow the expanding chest wall because their surfaces remain coupled to it through the thin fluid-filled pleural cavity. More negative pleural pressure increases the distending transpulmonary pressure; enlargement of the alveoli briefly makes alveolar pressure lower than atmospheric pressure, so air enters. At the end of inspiration, airflow stops when these last two pressures equalize. During quiet expiration, relaxation of the respiratory diaphragm and inspiratory intercostal muscles allows the stretched lungs and chest-wall system to recoil towards their resting balance. The domes rise, the ribs and sternum descend, and alveolar pressure becomes slightly positive relative to atmospheric pressure while air leaves. Active abdominal muscles and the interosseous portions of the internal intercostal muscles become important in forced expiration, rather than supplying the main force in quiet expiration.[ref]
22.3 The Process of Breathing. Mechanisms of Breathing and Pulmonary Ventilation; Figures 22.16–22.17
A simple mechanical demonstration uses two elastic bags in an airtight transparent vessel, connected through its stopper to the air, with a movable rubber sheet forming the base. Pulling the sheet down expands the vessel and inflates the bags; letting it return allows deflation. A side-tube manometer can show the associated pressure change. This illustrates expansion by lowering the floor, but its rigid vessel cannot demonstrate the movements of the ribs, the action of the intercostal muscles, or the real geometry of the pleural cavity. Observing lower-chest widening and abdominal displacement during quiet breathing supplies those missing observations without confusing the model with the anatomy.
Pleural pressure 2026-10-31
Positive-pressure ventilation 2026-11-05
Positive-pressure ventilation assists inhalation by raising pressure at the airway opening, driving gas into the lungs when that pressure exceeds alveolar pressure. It differs mechanically from spontaneous expansion of the thorax or negative-pressure ventilation, which lowers pressure around the lungs to produce the relevant distending difference. A ventilator's mask or tube supplies the applied airway connection.[ref]
The direction of flow still depends on a pressure difference, not on whether either pressure is above atmospheric. For a simple resistive path,with positive directed inward and the airway resistance. In an original example, pressures 10 and 8 cm HO and cm HO·s/L give inward flow 0.5 L/s. Pressures 8 and 10 cm HO give outward flow 0.5 L/s. Both pairs are positive relative to atmosphere.
This explains why positive end-expiratory pressure can coexist with exhalation. The applied pressure can remain above atmospheric while elastic recoil produces a still higher alveolar pressure and drives gas outward. At equal opening and alveolar pressures, the same ideal resistive model gives zero flow.
The resulting tidal volume also depends on respiratory-system compliance, airway resistance, duration and any active respiratory muscle effort. The no-flow interval used to interpret plateau airway pressure removes one resistive contribution only under the required measurement conditions.[ref] Positive pressure is therefore a mode of applying mechanical assistance, not a guarantee of a particular inward flow, volume or gas-exchange result.
Spirometry 2026-11-05
Spirometry measures the volume of air moved into or out of the respiratory system as a function of time. A spirometer records displaced gas directly or derives volume by integrating volumetric flow rate. The resulting trace describes gas movement at the mouth; it does not independently determine alveolar pressure, pleural pressure or lung compliance.[ref]
Standardization of Spirometry 2019 Update by Brian L. Graham and others. Introduction and Equipment: volume and flow measurements
Two important outputs of a forced expiration following maximal inspiration are forced vital capacity, the total volume expelled in that maneuver, and forced expiratory volume in one second, the volume expelled during its first second. Their ratio compares early emptying with the eventual expelled volume.[ref]
Standardization of Spirometry 2019 Update by Brian L. Graham and others. FVC Maneuver and definitions of FVC and FEV1
Equal final volumes can accompany substantially different emptying rates.An original mathematical example sets expelled volumewhere the numerical time and time constant are in seconds. Both and give an asymptotic expelled volume of 4 L. Their first-second volumes are instead L and L. The corresponding ratios to the limiting volume are approximately 0.632 and 0.393. These smooth toy curves explain the measurements; they are not measured traces or diagnostic reference limits.
Spirometry measures volume differences. The gas left after maximal expiration is residual lung volume, which has not crossed the instrument. Consequently a volume trace alone does not determine total lung capacity: different absolute starting volumes can produce the same expelled-volume trace. This is the absolute-volume ambiguity in spirometry.
Timing, effort, leaks and early termination can alter what a trace represents. Distinguish an instrumentally repeatable signal from a maneuver that answers the intended physiological question. The primary technical statement treats equipment quality assurance and maneuver acceptability separately.[ref]
Standardization of Spirometry 2019 Update by Brian L. Graham and others. Device Quality Assurance and FVC Maneuver acceptability sections
Static versus dynamic lung compliance 2026-11-05
Static versus dynamic lung compliance distinguishes an equilibrated elastic pressure-volume relation from a breathing-cycle measurement. Lung compliance uses transpulmonary pressure, the difference between alveolar pressure and pleural pressure. A static local value is after flow and relevant transients have subsided; a finite interval instead gives over a stated pressure range.
The timing convention for dynamic measurements must be explicit. Conventional dynamic lung compliance uses volume and transpulmonary-pressure changes between the zero-flow endpoints of a breath. It is not automatically the same quantity as a ratio using peak airway pressure. The respiratory-function treatise distinguishes these definitions.[ref]
Pulmonary Function Testing by Warren M. Gold and Laura L. Koth. Static and Dynamic Compliance: dynamic lung measurements use zero-flow endpoints of a tidal breath
An original one-compartment model isolates flow contamination. Relative to a chosen baseline, let mouth-to-pleural driving pressure satisfyWith stipulated , , L and , the pressure is cm water. Dividing volume by that pressure gives 0.10 rather than the actual elastic compliance 0.20. Setting flow to zero removes the resistive term in this ideal model. This is a demonstration of a flow-contaminated pressure ratio, not a redefinition of the zero-flow dynamic measurement.
Even zero flow at the mouth need not establish full internal equilibration. Unequal filling times in apparent lung compliance can be illustrated with two independent elastic compartments receiving the same imposed pressure step. If their time constants are , thenFor in compliance units and s, s, the effective response after 0.5 s is about 0.114, whereas its long-time limit is 0.20. The shorter observation captures less filling of the slow compartment. These are constructed passive model values, not clinical reference ranges.
A primary 2022 pilot study compared four static and dynamic calculation methods during passive mechanical ventilation. Its methods distinguish plateau pressure, pressure at zero flow, constant-flow slopes and expiration-derived estimates.[ref]
Static and Dynamic Measurements of Compliance and Driving Pressure: A Pilot Study by Pierre Tawfik and others. Figure 1 and Introduction, Methods A–D; Materials and Methods, passive ventilation
Those airway-pressure measurements describe the respiratory system, including the chest wall. Separating lung mechanics requires an additional internal pressure reference, such as the esophageal-balloon approach used in the 1959 experiment.[ref]Airway Resistance and Lung Compliance by Ole Storstein and others. Methods, intra-esophageal balloon and differential pressure recording