A. Buursma 2026-11-05
A. Buursma collaborated with W. G. Zijlstra and W. P. Meeuwsen-van der Roest on experimental reference spectra for hemoglobin derivatives. Their 1991 paper measured molar absorption coefficients for fetal and adult forms of the oxygenated, deoxygenated, carbon-monoxide-bound and oxidized derivatives.[ref]
Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin by W. G. Zijlstra and others. Byline and Abstract
Such measurements support quantitative optical analysis of mixtures, including co-oximetry.The measurement problem can be expressed through the Beer-Lambert law. For a transparent mixture with a known path length , a simple linear model at wavelength isHere is absorbance, are component concentrations and are reference molar absorption coefficients. The reference experiment provides coefficients; the later mixture experiment measures the . These are distinct roles in the analysis.
An original example illustrates why multiple independent readings are useful. With normalized coefficients, let and . Measurements imply . With only the first reading, both and fit. Additional readings help only when their coefficient rows supply linear independence; repeating the same equation more precisely does not identify both concentrations.
The example explains the importance of reliable reference coefficients without attributing an unreported individual laboratory task to A. Buursma. The verified scientific contribution is coauthorship of the experimental spectral dataset.
Arterial blood gas analysis 2026-11-05
Arterial blood gas analysis applies blood gas analysis to a sample taken from an artery, providing a time-specific observation of systemic arterial gas tensions and acid-base state. Arterial oxygen partial pressure concerns oxygen tension, while arterial carbon-dioxide partial pressure and pH provide different information. The arterial qualifier is a sampling condition, not merely a label attached to any blood result.[ref]
Arterial Blood Gas (ABG) Test by MedlinePlus. What is an arterial blood gas test? measured quantities
The measurement chain includes obtaining an identified sample, preserving its gas composition and analyzing it with a checked instrument. Air contamination, fluid contamination and analysis delay can change the specimen after it leaves the circulation. A reproducible analyzer reading cannot repair a specimen that no longer represents the intended physiological state.[ref]
Equal oxygen tension need not mean equal arterial oxygen content. An original illustrative carrier model writeswhere is hemoglobin concentration, its oxygen-bound fraction and the carrier capacity per gram. Choose mL/g, mL/(dL·mmHg), and mmHg. Samples with and g/dL then contain about 20.00 and 10.15 mL O/dL, respectively. They have identical stipulated tension and saturation but very different oxygen content. The constants define the illustration; these are not patient results or diagnostic thresholds.
Interpretation also depends on the conditions when the sample was obtained. Inspired oxygen fraction, ventilator settings when applicable and sampling time help distinguish a changed respiratory state from a changed measurement condition. A later pulse oximetry reading is neither a simultaneous sample nor a measurement of carbon-dioxide tension.
Arterial blood gas analysis supplies evidence about gas transport and acid-base balance, but no individual number alone identifies the entire cause of an abnormal result. Measured co-oximetry and calculated oxygen saturation should also be distinguished when unusual hemoglobin species matter.[ref]
Blood gas analysis 2026-11-05
Blood gas analysis characterizes a specified blood specimen using gas tensions and acid-base measurements. Core measured quantities include oxygen and carbon dioxide partial pressures and pH; instruments can also report derived quantities. Co-oximetry supplies related information about hemoglobin species by optical measurement rather than inferring all of them from gas tensions.[ref]
AARC Clinical Practice Guideline: Blood Gas Analysis and Hemoximetry: 2013 by Michael D. Davis and others. Section 1.0, Description, measured and derived variables
The specimen's source is part of the result. An arterial sample addresses a different compartment from a peripheral venous sample, where tissues have already exchanged gases with the blood. The sample is analyzed outside the circulation, so collection time, inspired gas, temperature convention and sample handling affect interpretation. The guideline identifies contamination, delay and temperature conventions as distinct limitations.[ref]
A calculation can look like an additional measurement. For an illustrative bicarbonate equilibrium approximation at a stated temperature, the Henderson-Hasselbalch equation can be writtenwith bicarbonate concentration in mmol/L and carbon-dioxide partial pressure in mmHg. The numerical constants specify this approximation; they are not universal across all temperatures and media. Rearranging givesFor stipulated pH 7.40 and mmHg, the calculated value is about 23.94 mmol/L. Holding pH fixed and changing that partial pressure to 60 mmHg raises the calculation to about 35.91 mmol/L. Those derived concentrations do not constitute a third independent observation validating the two inputs.
A sample-specific gas partial pressure also differs from total gas content. Much of transported oxygen is bound to hemoglobin, while measured oxygen tension concerns dissolved oxygen. A panel should therefore identify which values were directly measured, which were computed and which specimen supplied them. Agreement among quantities calculated from one another is internal arithmetic consistency, not independent physiological confirmation.
Carboxyhemoglobin 2026-11-05
Carboxyhemoglobin is hemoglobin with carbon monoxide bound at a haem site. Carbon monoxide competes with oxygen for binding; a CO-occupied site is unavailable for oxygen while CO remains bound.[ref]
Hemoglobin by Radiometer. Additional manufacturer parameter page, Carboxyhemoglobin, What is COHb?
This differs from methemoglobin, where the defining change is oxidation of haem iron. Binding carbon monoxide is not itself the same operation as converting an available ferrous site into a ferric site.An original independent-site equilibrium model illustrates competition. Let oxygen and carbon-monoxide binding weights be and , with values defined on the same pressure basis. For empty, oxygen-bound and CO-bound states,With stipulated and no carbon monoxide, the oxygen-bound fraction is . Adding a competing weight changes it to and gives CO the same fraction. These weights are hypothetical; this model omits the cooperative and conformational behavior of real hemoglobin.
The calculation explains why knowing the oxygen partial pressure alone does not identify the amount bound to hemoglobin. Available-site count, competing ligands and binding properties also enter. Even a high occupied fraction among the remaining oxygen-capable sites does not imply that all haem sites are oxygenated.
Co-oximetry separates carboxyhemoglobin from other derivatives using their spectral responses. The 1991 fetal/adult hemoglobin study quantified derivative absorptivities and emphasized that even small spectral differences affect multicomponent analysis.[ref]
Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin by W. G. Zijlstra and others. Abstract
Identifying a derivative therefore requires a suitable measurement model; total hemoglobin concentration alone leaves its ligand composition unresolved. Functional oxygen saturation 2026-11-05
Functional oxygen saturation is the fraction of the sum of oxyhemoglobin and deoxyhemoglobin that is present as oxyhemoglobin:Here and are the corresponding concentrations in the same blood sample. Carboxyhemoglobin and methemoglobin are excluded from this denominator. Fractional oxyhemoglobin instead divides by the total concentration of measured hemoglobin species.[ref]
Oxyhemoglobin by Radiometer. FO2Hb versus oxygen saturation (sO2)
A constructed example shows what the exclusion does. If a sample contains 90 concentration units of oxyhemoglobin, 10 of deoxyhemoglobin and 25 of carboxyhemoglobin, its functional oxygen saturation is . Its fractional oxyhemoglobin is . Both values describe the same mixture, with different reference populations.
If denotes the fraction of total hemoglobin excluded from the functional denominator, thenThus knowing only does not determine the oxygenated fraction of total hemoglobin unless is also known or negligible. In the example, and .
The distinction has an experimental consequence for co-oximetry. Optical analysis must distinguish the absorbing hemoglobin derivatives before calculating the desired ratio; changing the denominator of the displayed result does not change the blood sample. Total oxygen content additionally requires hemoglobin concentration and the dissolved oxygen contribution. Hemoglobin-concentration and saturation separation therefore remains necessary even when functional oxygen saturation has been measured accurately.
Isosbestic point 2026-11-05
An isosbestic point is a wavelength, wavenumber or frequency at which a sample's total absorbance remains unchanged during a specified chemical or physical transformation. It is a property of the observed spectra and transformation, rather than proof of a particular reaction mechanism.[ref]
IUPAC Compendium of Chemical Terminology, fifth edition (Gold Book) by International Union of Pure and Applied Chemistry. isosbestic point, I03310, definition and stoichiometric qualifications
For a simple conversion between two species and , assume additive Beer-Lambert law response, fixed optical path length and constant total concentration . ThenAt a wavelength where ,independent of the conversion fraction. For example, stipulated values , and give constant absorbance .
An isosbestic point does not establish that exactly two species exist. Three species could all have the same absorption coefficient at ; every mixture with the same total concentration would then give the same absorbance there. More generally, constant stoichiometric combinations of coefficients can produce an invariant signal. Conversely, dilution, changing path length, background absorption or changing spectra can remove a crossing without demonstrating a new reaction intermediate.
In an experimental test, record multiple full spectra while controlling total concentration, path length and background. Inspect the stability of the common intersection relative to measurement uncertainty, rather than selecting one pair of curves that happens to cross. The invariant wavelength alone cannot determine composition in the two-species model: its response has deliberately lost sensitivity to the conversion fraction. Measurements away from that wavelength carry the separating information, a distinction also useful in co-oximetry.
Methemoglobin 2026-11-05
Methemoglobin is a hemoglobin derivative with oxidized ferric iron at a haem site. The ferric site cannot perform the usual reversible oxygen binding of an available ferrous site. The distinction is one of oxidation state, not merely whether an oxygen molecule currently occupies the site.[ref]
Hemoglobin by Radiometer. Additional manufacturer parameter page, Methemoglobin, What is MetHb?
An adult hemoglobin tetramer contains four haem sites, so site occupancy and the state of the whole tetramer must not be silently conflated.Deoxyhemoglobin retains the capacity to bind oxygen at its available sites. A ferric site in methemoglobin instead requires chemical reduction to restore the corresponding ferrous state. Formation of oxyhemoglobin is therefore different from accumulating ferric haem. Changes in the surrounding protein also influence the properties of remaining sites; a site-count calculation alone is not a complete cooperative-binding model.
For an original capacity calculation, suppose a sample contains 100 micromoles of haem sites, with 20 micromoles ferric and the other 80 capable of reversible oxygen binding. The latter can bind at most 80 micromoles of oxygen in a one-molecule-per-site model. If 60 micromoles are oxygenated, the oxygenated fraction of all haem sites is , while the occupied fraction of the stipulated capable sites is . A denominator change produces different percentages without changing the sample.
Experimentally, spectrophotometry distinguishes derivatives through calibrated absorption spectra. A 1991 primary study measured human fetal and adult oxyhemoglobin, unoxygenated hemoglobin, carboxyhemoglobin and methemoglobin spectra over 450–1000 nm.[ref]
Absorption spectra of human fetal and adult oxyhemoglobin, de-oxyhemoglobin, carboxyhemoglobin, and methemoglobin by W. G. Zijlstra and others. Abstract, measured millimolar absorptivities
The relevant inverse problem has the formwhere path length , derivative concentrations and their calibrated absorption coefficients determine the modeled absorbance. Co-oximetry uses multiple spectral measurements to estimate such derivative fractions. A single colour observation cannot generally separate several unknown components. Pulse oximetry 2026-11-05
Pulse oximetry estimates arterial oxygen saturation noninvasively from wavelength-dependent optical signals that vary with the arterial pulse. A typical sensor combines red and infrared light-emitting diodes with a photodetector. The pulsatile component helps distinguish changing arterial blood contribution from relatively steady tissue and blood contributions; it is not a direct measurement of arterial oxygen partial pressure.
A primary instrument-design note implements 660 nm and 940 nm illumination, a shared photodetector and alternating channel acquisition. It separates pulsatile and steady components before estimating saturation.[ref]
Pulse Oximeter Fundamentals and Design by Santiago Lopez. Sections 2.2 and 3.1.2, printed pp. 3–5 and 7, optical channels and acquisition
These wavelengths and electronics describe that implementation rather than every possible sensor.Writing pulsatile and steady signal magnitudes as AC and DC, a common normalized optical quantity isFor original channel values AC, DC, AC and DC, this ratio is 0.5. Halving both red-channel magnitudes leaves the ratio unchanged. This illustrates intensity normalization; it does not determine a saturation without a validated relation between and saturation.
That relation is experimentally checked against an independent reference. The FDA's technical guidance describes comparing oximeter readings with simultaneous co-oximetry of arterial samples during controlled inspired-oxygen changes in volunteers, and reporting an accuracy statistic.[ref]
Pulse Oximeters: Premarket Notification Submissions by the FDA. Clinical Performance, SpO2 Accuracy: reference arterial sampling, inspired mixtures and paired readings
Such comparisons test the whole sensor–algorithm measurement chain, rather than only the arithmetic of a two-channel ratio.One summary is the root mean square differenceFor invented paired errors percentage points, percentage points even though mean error is zero. A small mean bias does not guarantee small individual errors.
Motion, weak pulsatile signals and additional absorbing hemoglobin species can invalidate assumptions behind a simple estimate. Pulse oximetry does not measure pH, carbon-dioxide tension or hemoglobin concentration. A stable displayed percentage therefore cannot establish every aspect of ventilation, oxygen content or gas exchange.
W. P. Meeuwsen-van der Roest 2026-11-05
W. P. Meeuwsen-van der Roest coauthored a 1991 experimental study with W. G. Zijlstra and A. Buursma measuring the molar absorption coefficient of four derivatives of fetal hemoglobin and adult hemoglobin across 450–1000 nm. The work supplied spectral reference measurements relevant to separating oxyhemoglobin, deoxyhemoglobin, carboxyhemoglobin and methemoglobin by optical analysis.[ref]
The scientific point is that similar-looking spectra need not be interchangeable reference data. In co-oximetry, a model based on the Beer-Lambert law combines reference spectra to infer component concentrations. A small reference mismatch can alter the inferred mixture even when the measured absorbance is precise.
An original two-component example explains the sensitivity. Suppose two synthetic readings satisfyThen . If , an error of in changes the inferred by in the model's concentration units. Near-similarity of reference columns can therefore amplify small errors. These invented coefficients illustrate the inverse problem; they are not the paper's measured hemoglobin spectra.
