The finesse of a Fabry–Pérot interferometer is its free spectral range divided by a transmission peak's full width at half maximum. Large finesse means a narrow accepted frequency interval compared with the spacing between repeated orders. Here half maximum means of the ideal unit peak, with zero detector background subtracted; it does not mean halfway between the transmission minimum and maximum.
Write for the coefficient multiplying in Fabry–Pérot transmission from multiple reflections. This coefficient is sometimes called the coefficient of finesse; it is not the finesse . About a resonance , put . Solving gives
and, since one free spectral range advances the wave phase by ,
These are exact for the ideal nondispersive model with constant and linear frequency dependence of . The inverse sine requires , or . At equality, neighbouring peaks merely touch half maximum at their shared minima: the nominal width is one full free spectral range and . Isolated half-maximum peaks require the strict inequality. Below this threshold the transmission never falls to , so this peak-width definition supplies no finesse; at the transmission is flat.
For large reflectivity, the small-angle approximation gives
This is a high-reflectivity approximation, not an identity for every . The original dimensionless examples below have exact full width at half maximum values obtained from the inverse sine formula.
Mirror intensity reflectivity Exact Exact finesseApproximate finesse
Figure 1.
Increasing mirror reflectivity narrows Fabry–Pérot interferometer transmission peaks
. Original stipulated dimensionless model for , with and modulo . All curves peak at unity and repeat after . The second panel enlarges the central resonance; coloured segments at mark each exact full width at half maximum. The inset compares exact finesse with the high-reflectivity approximation. These are calculations, not measured scans. Details.
The beauty of the narrowing is visible directly in the geometric series: large preserves contributions from many round trips, so a small phase difference accumulates across many terms and spoils their addition. For the stipulated vacuum gap, gives an ideal full width at half maximum of about . This instrumental width is not a measured spectral line width. A real scan also depends on the source profile, accepted angles, mirror shape, alignment, and scan stability. The University of Washington handout separately discusses mirror-reflectivity, mirror-flatness, and detector-pinhole contributions; increasing alone cannot remove the others.[ref]
For an ideal air-gap Fabry–Pérot etalon at normal incidence, the free spectral range is and the peak full width at half maximum is . To separate a pair of lines spaced by by at least peak widths, require . Putting both lines inside one unambiguous free spectral range additionally requires , so . In terms of their vacuum wavelengths, .
The strict order condition needs a margin if tails from adjacent periodic peaks are to be negligible. If every neighbouring peak in a scanned two-line periodic spectrum must be separated by widths, the stronger condition is . Ordinary resolving power alone does not prevent ambiguous repeated orders. A fixed etalon's transmission as a function of wavelength is a periodic-in-frequency filter; it is not automatically a plot of the incident source's spectral lines.