Fabry–Pérot interferometer Created 2026-10-05 Updated 2026-10-14
A Fabry–Pérot interferometer uses repeated specular reflections between two nearly parallel, partially transmitting surfaces. Its transmitted light is the sum of contributions that have made zero, one, two, or more additional round trips. Their complex amplitudes reinforce only near particular wave phases, producing narrow transmission peaks that select wavelength intervals. Scanning the spacing or angle of incidence can build an imaging spectral cube: an image at each selected wavelength.
The surprising feature is that highly reflecting mirrors can together transmit all the incident power at resonance in the ideal symmetric, lossless model. Each individual contribution is weak, but many contributions arrive in phase. Between resonances the same contributions cancel strongly. Increasing mirror reflectivity sharpens the peaks instead of simply suppressing every transmitted signal; finesse quantifies that selectivity. An optical spectrum with nearby spectral lines can therefore produce distinguishable peaks, provided their separation is large enough relative to the instrument's peak width and their orders are identified.
Charles Fabry and Alfred Pérot developed the instrument together; Optica dates its development from 1896 to 1898.[ref]
Charles Fabry, Optica History biography. Charles Fabry biography, paragraphs on collaboration with Alfred Pérot and instrument development
Finesse 2026-10-14
The finesse of a Fabry–Pérot interferometer is its free spectral range divided by a transmission peak's full width at half maximum. Large finesse means a narrow accepted frequency interval compared with the spacing between repeated orders. Here half maximum means of the ideal unit peak, with zero detector background subtracted; it does not mean halfway between the transmission minimum and maximum.
Write for the coefficient multiplying in Fabry–Pérot transmission from multiple reflections. This coefficient is sometimes called the coefficient of finesse; it is not the finesse . About a resonance , put . Solving givesand, since one free spectral range advances the wave phase by ,These are exact for the ideal nondispersive model with constant and linear frequency dependence of . The inverse sine requires , or . At equality, neighbouring peaks merely touch half maximum at their shared minima: the nominal width is one full free spectral range and . Isolated half-maximum peaks require the strict inequality. Below this threshold the transmission never falls to , so this peak-width definition supplies no finesse; at the transmission is flat.
For large reflectivity, the small-angle approximation givesThis is a high-reflectivity approximation, not an identity for every . The original dimensionless examples below have exact full width at half maximum values obtained from the inverse sine formula.
Increasing mirror reflectivity narrows Fabry–Pérot interferometer transmission peaks
. Original stipulated dimensionless model for , with and modulo . All curves peak at unity and repeat after . The second panel enlarges the central resonance; coloured segments at mark each exact full width at half maximum. The inset compares exact finesse with the high-reflectivity approximation. These are calculations, not measured scans. Details.The beauty of the narrowing is visible directly in the geometric series: large preserves contributions from many round trips, so a small phase difference accumulates across many terms and spoils their addition. For the stipulated vacuum gap, gives an ideal full width at half maximum of about . This instrumental width is not a measured spectral line width. A real scan also depends on the source profile, accepted angles, mirror shape, alignment, and scan stability. The University of Washington handout separately discusses mirror-reflectivity, mirror-flatness, and detector-pinhole contributions; increasing alone cannot remove the others.[ref]
Fabry-Perot Interferometer, University of Washington Physics 331. Finesse discussion, equations (5)–(8), and §2.3 Determine the Finesse
Scanning Fabry–Pérot measurement 2026-10-14
An actual teaching experiment in LM-0300 Fabry Perot Spectrum Analyser by Walter Luhs, §§3.1.1–3.1.2, prepares a green laser beam by aligning it near and far along a rail, then centring back-reflected light from adjustable curved mirrors. An actuator using piezoelectricity moves one mirror. A screen first displays interference fringes; a focusing optical lens and photodiode then replace the viewing arrangement. The oscilloscope records the photodiode signal and the piezo drive reference on separate channels, using the latter as its trigger. Alignment and mirror separation are adjusted to suppress extra transverse normal modes; source temperature and drive electric current are controlled. Repeated peaks calibrate the scan using the appropriate free spectral range, and their spacing divided by their full width at half maximum estimates finesse.[ref]
This experiment uses a confocal curved-mirror cavity, whose repeated peak families can have spacing in air; it must not be calibrated blindly with the parallel-mirror . Extra transverse normal modes can mimic source features. A finite measured peak width bounds source sharpness only after instrumental broadening is considered; it does not determine the source's intrinsic width by itself.[ref]
LM-0300 Fabry Perot Spectrum Analyser by Walter Luhs. §3.1.2 p. 20: transverse modes, confocal calibration and source-linewidth inference
For an independent ideal scan calculation, hold a narrow laser wavelength fixed and vary the parallel-mirror gap at normal incidence. Advancing one resonance order requires . If changes uniformly over the measured interval, the time between repeated peaks divided by their full width at half maximum in time equals finesse. The cancellation of scan rate follows from : both peak spacing and peak width are divided by the same . It fails if the scan rate changes appreciably over that interval. A practical control in the University of Washington's motor-driven flat-mirror experiment is to check that successive peaks are equally spaced before using the time axis this way; its handout also warns that table strain and aggressive low-pass filtering alter the trace.[ref]