Absorbed dose 2026-10-20
Mean energy imparted by ionizing radiation per unit mass; its unit is the gray.
The idealized external photon estimate is approximately at one metre. First apply radioactive decay to the certified radioactive activity:
Treat the source as isotropic and small compared with the observation distance. Neglect photon self-absorption, capsule attenuation and attenuation in air. The metal and capsule stop the primary beta particles, so the direct external contribution is taken to be the two gamma rays; secondary bremsstrahlung is neglected. Their total energy per transformation is
The incident photon energy flow per area at distance is therefore
Interpret the given as the linear energy-absorption coefficient , rather than a total photon-removal coefficient. For an absorbing medium of mass density , a thin volume absorbs energy per time and has mass . Its absorbed dose rate is consequently
For the estimate assume a water-like and local deposition of the secondary charged-particle energy. At ,
For gamma rays, the radiation weighting factor is one, giving under these assumptions.[ref]
The phrase “equivalent absorbed dose” mixes two quantity names: absorbed dose is measured in gray, while equivalent dose is measured in sievert. Their rates are numerically equal here because the weighting factor is one. The estimate is a physical energy-deposition calculation; the assumed mass density is essential, and the stated inverse-length coefficient must not be used as if it already had area-per-mass units.
Sievert 2026-10-20
The SI unit used for equivalent dose and effective dose. Sharing dimensions with the gray does not make the quantities interchangeable.[ref]